Uninterruptable power supply (ups) actions – MTS Series 793 User Manual
Page 297
Description
Item
Select an action to occur in the event the UPS supplies power to the system and the
On UPS Delay time has expired.
On UPS Delay
Actions
Select an action in the event the UPS battery becomes low while the UPS supplies
power to the system.
On UPS Low Battery
Actions
Select an action in the event the UPS battery becomes low while line voltage supplies
power to the system.
UPS On-line Low
Battery Actions
Select an action in the event the external event associated with the UPS is detected.
UPS External Event
Actions
Uninterruptable Power Supply (UPS) Actions
UPS Action Settings
Action
Setting
No action occurs. This setting dims the indicator.
Disabled
Writes a message to the Message Log and the indicator switches from white
to red.
Indicate
Stops any program command, writes a message to the Message Log, turns
off pressure at the hydraulic service manifold (HSM), and switches the
indicator from green to red.
Station Power Off
For single station systems: If the “Last Off” function is enabled in the .hwi
file or equivalent jumpering, the hydraulic power unit (HPU) is turned off. If
the “Last Off” function is disabled, the HPU will remain on.
For multi station systems: If the “Last Off” function is enabled in the .hwi
file or equivalent jumpering, the hydraulic power unit (HPU) will be turned
off only if the interlock occurs on the last station running. If the “Last Off”
function is disabled, the HPU will remain on.
Stops any program command, writes a message to the Message Log,
clamps the servovalve (if enabled in the .hwi file), turns off pressure at the
Interlock
hydraulic service manifold (HSM), and switches the indicator from green to
red.
Note:
To resume testing, fix the condition that caused the interlock and
then reset the Interlock.
For single station systems: If the “Last Off” function is enabled in the .hwi
file or equivalent jumpering, the hydraulic power unit (HPU) is turned off. If
the “Last Off” function is disabled, the HPU will remain on.
For multi station systems: If the “Last Off” function is enabled in the .hwi
file or equivalent jumpering, the hydraulic power unit (HPU) will be turned
MTS Series 793 Control Software 297
Station Setup