Description

Item

Adjust for the required -mV/V compression sensitivity value.

Neg. Compression (mV/V
Pos Tension only)

Adjust for the required +mV/V tension sensitivity value.

Pos. Tension (mV/V Pos
Tension only)

Adjust for the required -mV/V tension sensitivity value.

Neg. Tension (mV/V Pos
Compression only)

Adjust for the required +mV/V compression value.

Pos. Compression (mV/V Pos
Compression only)

(Model 494.16/.25/.26 DUCs only) Sets the frequency of the excitation voltage
applied to the sensor.

Discrete Excitation

Sets the excitation applied to the sensor. The proper excitation is usually
specified by the sensor manufacturer.

Excitation (peak) for AC
Conditioners

Typically AC sensors require a 10 V AC excitation signal and DC sensors
require a 10 V DC signal.

Excitation (p-p) for DC
Conditioners

Adjusts the phase of the feedback signal based on the excitation signal.

Phase (AC only)

Optimal phase adjustments produce the maximum output signal from the
sensor. Phase values are usually between 30º–45º.

Specifies the amount of delta K applied to the sensor’s output signal.

Delta K

Equally negative and positive inputs cause asymmetry in the sensor’s output
signal. Delta K compensates for this asymmetry.

For mV/V Pos Tension and mV/V Pos Comp calibration types, Delta K is a
read-only display.

Adjusts the offset applied by the conditioner to the sensor’s output signal to
set this signal’s zero point.

Zero

Zero controls vary by conditioner type.

Zeroes the sensor signal to make the current output the new zero point.

Auto Zero (DC only)

Applies a fine zero offset to the sensor’s output signal.

Fine Zero (if available)

For Model 497.22 DC Conditioners—Provides either pre-amp offset (coarse
zero) or bridge balance control, depending on Model 497.22 DC jumpering.

Zero/Balance (DC only)

See the Model 497.22 Dual DC Conditioner manual for more information
about the required jumpering.

You cannot save a sensor file while the system is running.

Save

Saves settings from the Sensor, Calibration, and Shunt tabs to the current
sensor file.

Displays the Save Sensor File As window if no file has been created.

Restores the values on the Sensor, Calibration, and Shunt tabs from the
sensor file or Transducer ID.

Restore

MTS Series 793 Control Software 201

Station Setup