beautypg.com

How to set up null pacing compensators – MTS Series 793 User Manual

Page 247

background image

Description

Item

Sets the time allowed for feedback to come within tolerance.

Timeout

Sets the test station’s action when the Timeout is exceeded. The actions
available are identical to those available for Static Null Pacing.

Timeout Action

How to Set Up Null Pacing Compensators

If only static null or dynamic null pacing is required, set the Error Tolerance for the unused compensation
method to 100%.

1. Select an access level of Tuning.

2. On the Station Setup window navigation pane, locate and select the Channels control mode needing

compensation.

3.

In the Station Setup window, click

.

4. On the Compensators panel, click the Null Pacing tab.

Warning:

Changes in compensation values made with hydraulic pressure on can result in unexpected
actuator movement.

A moving actuator can injure anyone in its path.

Always clear the actuator area before changing compensation values.

5. On the Null Pacing tab, define Static Null Pacing values:

a) Set the Error Tolerance.

b) Set the Timeout values.

c) Select a Timeout Action.

6. On the Null Pacing tab, define Dynamic Null Pacing values:

a) Set the Error Tolerance.

b) Set the Timeout values.

c) Select a Timeout Action.

7. On the File menu, select Save Parameters to save the values.

MTS Series 793 Control Software 247

Station Setup