How to set up null pacing compensators – MTS Series 793 User Manual
Page 247
Description
Item
Sets the time allowed for feedback to come within tolerance.
Timeout
Sets the test station’s action when the Timeout is exceeded. The actions
available are identical to those available for Static Null Pacing.
Timeout Action
How to Set Up Null Pacing Compensators
If only static null or dynamic null pacing is required, set the Error Tolerance for the unused compensation
method to 100%.
1. Select an access level of Tuning.
2. On the Station Setup window navigation pane, locate and select the Channels control mode needing
compensation.
3.
In the Station Setup window, click
.
4. On the Compensators panel, click the Null Pacing tab.
Warning:
Changes in compensation values made with hydraulic pressure on can result in unexpected
actuator movement.
A moving actuator can injure anyone in its path.
Always clear the actuator area before changing compensation values.
5. On the Null Pacing tab, define Static Null Pacing values:
a) Set the Error Tolerance.
b) Set the Timeout values.
c) Select a Timeout Action.
6. On the Null Pacing tab, define Dynamic Null Pacing values:
a) Set the Error Tolerance.
b) Set the Timeout values.
c) Select a Timeout Action.
7. On the File menu, select Save Parameters to save the values.
MTS Series 793 Control Software 247
Station Setup