beautypg.com

About series 497 hardware limit detectors – MTS Series 793 User Manual

Page 290

background image

Item

Callout

Limit

1

Range

2

When you select a new range, review the limits you have defined for the selected signal to ensure that they
are appropriate for your new range.

About Series 497 Hardware Limit Detectors

For FlexTest IIm systems, a limit detector on each Series 497 Conditioner continuously compares its associated
transducer output value to preset upper and lower limit values set in the Hardware Limits tab.

When a conditioner’s upper or lower limit is exceeded, its limit detector is tripped, triggering an interlock.

Note:

Hardware limit detectors are supported as an option that must be enabled.

Hardware Limits Tab

Description

Item

Select to enable hardware interlocks when previously set conditioner limits or
excitation values are exceeded.

Enable Hardware
Interlocks (Limits and
Excitation)

If hardware interlocks are not enabled, interlock indicators will be gray and the
interlock will not occur.

Sets the upper limit that the conditioner’s feedback signal voltage must exceed
to trigger a hardware interlock.

Upper Limit

290 MTS Series 793 Control Software

Station Setup