MTS Series 793 User Manual
Page 310
3. Set the Integrator to manual mode.
a) In the Station Options window, click the Station View Options tab and set the Integrator Mode to Manual.
b) The Integrators control appears in the Station Manager Station Controls panel.
4. Save the parameter set to display the Integrators control panel each time the station is loaded.
How to Apply Hydraulics to a Station that Uses House Pressure
Use this procedure to apply hydraulic pressure to a station that does not control or monitor the hydraulic
pressure source.
Warning:
Unexpected actuator movement can occur when hydraulic pressure is applied to a station that
does not control or monitor hydraulic pressure.
Unexpected actuator movement can result in injury to personnel or damage to the equipment.
Follow these guidelines to minimize actuator movement when you apply hydraulic pressure to a
system that does not control or monitor hydraulic pressure: Ensure that the hydraulic pressure
is off before loading a station. Make sure that the mechanical and electrical valve balance is set
correctly. Before you apply hydraulic pressure to the system, use the Station Manager Manual
Command control to manually zero (or minimize) the error signal, disable the integrator, and
ensure that all program sources are off.
1. Set the Integrators Manual State setting to Disabled.
2. Use Manual Command to zero (or minimize) each error signal:
a) Use the Station Manager application to create a meter for the error signal for each channel.
b) In the Station Controls toolbar, click Manual Command.
c) Enable manual command and use the slider to zero (or minimize) the error signal for each channel.
d) Make sure that the output signal is stable.
3. Ensure that all program sources are off and click Reset/Override.
4. Apply hydraulic pressure to the system.
5. Set the Integrators Manual State setting to Enabled.
How to Remove Hydraulic Pressure from a Station that Uses House
Pressure
Use this procedure to remove hydraulic pressure to a station that does not control or monitor the hydraulic
pressure source.
310 MTS Series 793 Control Software
Using Station Manager to Run Tests