Fullscale tab, Channel input signals, How to access channel input settings – MTS Series 793 User Manual

Page 188

background image

Description

Item

Displays the current Setpoint value for each channel’s active control mode.
Setpoint values are not editable.

Setpoint

This group of indicators shows the status of each channel's Active Control Mode
signal. For example, if one of these signals becomes saturated, the Saturated

Status Indicators

[2] indicator turns red. Other signal status includes: Out of Range [1] (indicator
turns yellow), Invalid [3] (indicator turns red), or Disabled [4] (indicator turns
black).

Fullscale Tab

This tab displays editable minimum and maximum sensor ranges for each channel’s active control mode.

You can select a group of channels for display in the Channel List.

Access

Station Setup window > navigation pane > Channels > Fullscale tab

Fullscale Tab

Description

Item

Allows you to edit the displayed full-scale minimum and maximum range
for the sensor used in each channel’s active control mode.

Full scale

This requires the Calibration access level.

Channel Input Signals

How to Access Channel Input Settings

The Station Setup window provides two methods to access input settings.

Method 1

Click a control mode and then click the Channel Input Signals button to display the settings for that control
mode input.

188 MTS Series 793 Control Software

Station Setup