Aicx setup summary – MTS Series 793 User Manual

Page 265

background image

Description

Item

Warning:

Changes in compensation values made
with hydraulic pressure on can result in
unexpected actuator movement.

A moving actuator can injure anyone in
its path.

Always clear the actuator area before
changing compensation values.

Changing Anticipation or Impulse Response when
the filter is adapting automatically resets all filter
coefficients.

Clicking either Reset control (Fwd/Inv) with
compensation applied may produce an impulse to the
system. If this is undesirable, stop the function
generator before resetting.

Manages coefficients used in the compensation filter
and saved in the parameter set.

AICX Coefficients Reference

Coefficients do not display and do not change AICX
tab values.

Restores the saved AICX filter coefficients for the
selected channel.

Restore

Updates the coefficient references for the active
channel.

Update

To save the coefficient references, click Update and
then save the parameter set.

Enabled—Automatically downloads filter coefficients
when the current parameter set is saved and restored.

Auto-Restore

Disabled—Does not automatically download filter
coefficients. To restore filter coefficients, you must
click the Restore button.

AICX Setup Summary

1. Access the compensator.

2. Define AICX values.

3. Set up and enable limits to protect yourself and your equipment.

4. Install a dummy specimen.

MTS Series 793 Control Software 265

Station Setup