About electrical zero – MTS Series 793 User Manual

Page 209

background image

Item

Callout

Digital Signal

7

Offset

8

The offset applies to the feedback signal. You specify the offset in
engineering units.

Electrical zero considerations

Consider the following before shifting the electrical zero:

Unlike the Calibration tab’s conditioner Zero controls, electrical zero values are not used in the delta K
calculation.

The shifted electrical zero position is not a calibrated zero position.

You can shift the conditioner’s electrical zero to any position within the sensor’s full-scale capacity, but
your usable range is limited by the sensor’s calibrated full-scale capacity.

Linearization data

Using linearization data requires specific conditioner zeroing practices. Ensure that the Electrical Zero control
on the Offset/Zero tab of the Inputs panel is set to zero. Also, ensure that the Electrical Zero Lock box is
selected. Leaving electrical zero unlocked will invalidate data collected during linearization routines.

About Electrical Zero

Use of Electrical Zero with Gain/Linearization type calibrations

When you perform a Gain/Linearization type calibration on a Model 493.25 DUC in the linearization mode,
you must ensure that the controls pertaining to electrical zero are set properly.

Specifically, you must set:

The Electrical Zero control on the Offset/Zero tab of the Inputs panel to zero, and

The Electrical Zero Lock box to enabled.

The state of the Electrical Zero control and the Electrical Zero Lock box is saved in the station parameter set.

Electrical zero messages

If these conditions are not present when you attempt to load or save a parameter set, the application will
display one of the following messages.

Message

Message Location

Clearing the Electrical Zero Lock box on a Model 493.25 conditioner that
is using full-range calibration is not supported.

Message Log

Modifying Electrical Zero on a Model 493.25 conditioner that is using
full-range calibration is not supported.

Message Log

There are one or more Model 493.25 conditioners in which Electrical Zero
has been modified, and a full-range calibration is defined. This has the
potential to affect calibrations.

Dialog Box

MTS Series 793 Control Software 209

Station Setup