beautypg.com

How to adjust program output signals – MTS Series 793 User Manual

Page 230

background image

External programmers can:

Use the Station Manager application’s advanced programming resources.

Accommodate sensors that require special conditioning.

You must use the Station Builder application to create a Program Only, Program with Feedback, Program
and Control, or Command Plus Error program output channel using an analog output before the Station
Manager application can output a program to an external programmer.

How to Adjust Program Output Signals

1. On the Station Manager window toolbar, select an access level of Tuning.

2. On the Station Manager window Display menu, select Station Setup.

3. On the Station Setup window navigation pane, select the channel you want to adjust.

4.

On the Station Setup window, click

(channel drive).

Warning:

Changes in drive values made with hydraulic pressure on can result in unexpected actuator
movement.

A moving actuator can injure anyone in its path.

Always clear the actuator area before changing drive values.

5. On the Drive panel, make the following settings:

Program Output Settings

Settings

Channel Type

Set the Polarity, Valve Balance, Dither Amplitude, and Dither
Frequency.

Program and Control

Set Polarity, Gain, and Offset.

Program w/Feedback

Set the Polarity, Gain, and Offset and program output Fullscale
Min/Max values.

Program Only

230 MTS Series 793 Control Software

Station Setup