beautypg.com

Peak/valley phase compensators properties, How to set up peak/valley phase (pvp) compensators – MTS Series 793 User Manual

Page 271

background image

Item

Callout

Peak/Valley Compensator

3

PVP Compensator

4

Actuator

5

Peak/Valley Phase Compensators Properties

PVP Tab Settings

Description

Item

Sets how quickly compensation values get applied to converge the
feedback and command signal amplitudes.

Convergence Rate

Sets how much the feedback signal must change, as a percentage of its
peak-to-peak value, before a peak or valley is detected.

Sensitivity

Hold—Continues to use current values but stops adapting new values.

Adaptation State

Resume—Adapts and continuously updates compensation values.

Clears acquired compensation values and restarts the adaptation process.

Reset

Clicking Reset with compensation applied may produce a spike in the
feedback signal. Read the Warning that follows this table.

Warning:

Clicking Reset (or Reset All) in this tab with hydraulic pressure on can result in unexpected
actuator movement.

A moving actuator can injure anyone in its path.

Always clear the actuator area before clicking Reset.

How to Set Up Peak/Valley Phase (PVP) Compensators

1. Select an access level of Tuning.

2. On the Station Setup window’s navigation pane, locate and select the Channels control mode to receive

PVP.

3.

In the Station Setup window, click

4. On the Compensators panel, click the PVP tab.

MTS Series 793 Control Software 271

Station Setup