Operating Instructions

M6200E Rev: A1

99

6.9.9. Flow Calibration

The flow calibration allows the user to adjust the values of the sample flow rates as they are

displayed on the front panel and reported through COM ports to match the actual flow rate
measured at the sample inlet. This does not change the hardware measurement of the flow

sensors, only the software calculated values.

To carry out this adjustment, connect an external, sufficiently accurate flow meter to the sample

inlet (see Section 11.5.2 for more details). Once the flow meter is attached and is measuring actual
gas flow, press:

Exit at

any time
to return

to main

the

SETUP

menu

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

Repeat Pressing NEXT until . .

.

DIAG

FLOW CALIBRATION

PREV NEXT

ENTR

EXIT

DIAG FCAL

ACTUAL FLOW: 607 CC / M


0 6 0 7 ENTR EXIT

ENTR accepts the

new value

EXIT ignores the new

value

The value displayed is the

current output of the UV

source reference detector

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT