Operating Instructions

M6200E Rev: A1

89

To select manual output calibration for a particular channel, activate the ANALOG I/O
CONFIGURATION
MENU (see Section 6.9.1), then press:

DIAG AIO AOUT AUTO CAL: ON

ON ENTR EXIT

Exit to return to

the main

sample display

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AOUTS CALIBRATED:

NO

< SET

SET>

CAL

EXIT

DIAG AIO

CONC_OUT_2 RANGE: 5V

SET>

EDIT

EXIT

Toggles the

auto cal mode

ON/ OFF for

this analog

output channel

only.

DIAG AIO

CONC_OUT_2:5V, CAL

< SET SET>

EDIT

EXIT

Press SET> to select the analog output channel to

be configured. Then press EDIT to continue

ENTR accepts the new setting

and returns to the previous menu.

EXIT ignores the new setting and

returns to the previous menu.

DIAG AIO

CONC_OUT_2 REC OFS: 0 mV

< SET

SET>

EDIT

EXIT

DIAG AIO

CONC_OUT_2 AUTO CAL: ON

< SET

SET> EDIT

EXIT

Now the analog output channels should either be automatically calibrated or they should be set to
manual calibration, which is described next.

6.9.4.3. Manual Analog Output Calibration and Voltage Adjustment

For highest accuracy, the voltages of the analog outputs can be manually calibrated. Calibration is
done through the instrument software with a voltmeter connected across the output terminals
(Figure 6-5). Adjustments are made using the front panel keys by setting the zero-point first and

then the span-point (Table 6-12).

The software allows this adjustment to be made in 100, 10 or 1 count increments.

Table 6-12: Voltage Tolerances for Analog Output Calibration

Full Scale

Zero Tolerance

Span Voltage

Span Tolerance

0.1 VDC

±0.0005V

90 mV

±0.001V

1 VDC

±0.001V

900 mV

±0.001V

5 VDC

±0.002V

4500 mV

±0.003V

10 VDC

±0.004V

4500 mV

±0.006V

NOTE

Outputs configured for 0.1V full scale should always be calibrated manually