Model 6200E Instruction Manual

Operating Instructions

M6200E Rev: A1

87

To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION MENU (see
Section 6.9.1), then press:



DIAG AIO

AUTO

CALIBRATING CONC_OUT_1

AUTO

CALIBRATING CONC_OUT_2

AUTO CALIBRATING TEST_OUTPUT

Exit at any time

to return to the

main DIAG

menu

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AOUTS CALIBRATED:

NO

< SET

SET>

CAL

EXIT

If AutoCal has been

turned off for any

channel, the message

for that channel will be

similar to:

NOT AUTO CAL

CONC_OUT_1

Exit to return to

the I/O

configuration

menu

STARTING FROM DIAGNOSTIC MENU

(see Section 6.9.1)

If any of the channels have

not been calibrated this

message will read NO.

DIAG AIO AOUTS CALIBRATED:

YES

< SET

SET> CAL

EXIT

To automatically calibrate an single analog channel, activate the ANALOG I/O
CONFIGURATION
MENU (see Section 6.9.1), then press: