Model 6200E Instruction Manual

Operating Instructions

M6200E Rev: A1

103

6.10.2. COM Port Default Settings

As received from the factory, the analyzer is set up to emulate a DCE or modem, with pin 3 of the

DB-9 connector designated for receiving data and pin 2 designated for sending data.

COM1: RS-232 (fixed), DB-9 male connector.

o

Baud rate: 19200 bits per second (baud).

o

Data Bits: 8 data bits with 1 stop bit.

o

Parity: None.

COM2: RS-232 (configurable), DB-9 female connector.

o

Baud rate: 115000 bits per second (baud).

o

Data Bits: 8 data bits with 1 stop bit.

o

Parity: None.

CAUTION

Cables that appear to be compatible because of matching connectors may incorporate

internal wiring that make the link inoperable. Check cables acquired from sources other

than Teledyne Analytical Instruments for pin assignments before using.

6.10.3. RS-232 COM Port Cable Connections

In its default configuration, the M6200E analyzer has two available RS-232 Com ports accessible

via 2 DB-9 connectors on the back panel of the instrument. The COM1 connector is a male DB-9
connector and the COM2 is a female DB9 connector.

Male DB-9 (RS-232)

(As seen from outside analyzer)










(DTE mode)













(DCE mode)

1

2 3

4

5

6

7 8

9

RXD

GND

TXD

CTS

RTS

1

2 3

4

5

6

7 8

9

TXD

GND

RXD

RTS

CTS

Female DB-9 (COM2)

(As seen from outside analyzer)










(DTE mode)











1

2

3

4

5

6

7

8

9

RXD

GND

TXD

CTS

RTS

Figure 6-7: Back Panel connector Pin-Outs for COM1 & COM2 in RS-232 mode.