Figure 6-14: default idas channels setup – Teledyne 6200E - Sulfides Analyzer User Manual
Page 124
Operating Instructions
Model 6200E Instruction Manual
124
M6200E Rev: A1
LIST OF PARAMETERS
NAME: PNUMTC
EVENT: ATIMER
REPORT PERIOD: 001:00:00
NO. OF RECORDS: 360
RS-232 REPORT: OFF
CHANNEL ENABLED: ON
CAL HOLD OFF: OFF
NAME: CALDAT
EVENT: SLPCHG
NO. OF RECORDS:200
RS-232 REPORT: OFF
CHANNEL ENABLED: ON
CAL HOLD OFF: OFF
PARAMETER: SMPLFLW
MODE: AVG
PRECISION: 1
STORE NUM SAMPLES OFF
PARAMETER: SMPLPRS
MODE: AVG
PRECISION: 1
STORE NUM SAMPLES OFF
PARAMETER: S2OFS1
MODE: INST
PRECISION: 1
STORE NUM SAMPLES OFF
PARAMETER: S2SLP1
MODE: INST
PRECISION:3
STORE NUM SAMPLES OFF
PARAMETER: S2ZSC1
MODE: INST
PRECISION: 1
STORE NUM SAMPLES OFF
PARAMETER: S2CONC1
MODE: AVG
PRECISION: 1
STORE NUM SAMPLES OFF
PARAMETER: S2CONC1
MODE: AVG
PRECISION: 1
STORE NUM SAMPLES OFF
NAME: CONC
EVENT: ATIMER
REPORT PERIOD: 000:00:05
NO. OF RECORDS: 4032
RS-232 REPORT: OFF
CHANNEL ENABLED: ON
CAL HOLD OFF: ON
LIST OF CHANNELS
Figure 6-14: Default iDAS Channels Setup
These default Data Channels can be used as they are, or they can be customized from the front
panel to fit a specific application. They can also be deleted to make room for custom user-
programmed Data Channels.
Appendix A-5 lists the firmware-specific iDAS configuration in plain-text format. This text file can
either be loaded into APICOM and then modified and uploaded to the instrument or can be copied
and pasted into a terminal program to be sent to the analyzer.