Model 6200E Instruction Manual

TROUBLESHOOTING & REPAIR

M6200E Rev: A1

241

Table 11-1: Warning Messages - Indicated Failures

WARNING

MESSAGE

FAULT CONDITION

POSSIBLE CAUSES

ANALOG CAL

WARNING

The instruments A/D

circuitry or one of its

analog outputs is not

calibrated

A parameter for one of the analog outputs has been changed and the

calibration routine was not re-run

A/D circuitry failure on motherboard

Other motherboard electro9nic failure

BOX TEMP
WARNING

Box Temp is < 5

°

C or

> 48

°

C.

NOTE: Box temperature typically runs ~7

o

c warmer than ambient

temperature.

Poor/blocked ventilation to the analyzer.

Stopped exhaust-fan

Ambient temperature outside of specified range

CANNOT DYN

SPAN

Dynamic Span

operation failed

Measured concentration value is too high or low.

Concentration slope value to high or too low

CANNOT DYN

ZERO

Dynamic Zero

operation failed

Measured concentration value is too high.

Concentration offset value to high.

CONFIG

INITIALIZED

Configuration and

Calibration data reset

to original Factory

state.

Failed disk on chip

User erased data

CONV TEMP

WARNING

The temperature of the

H

2

S Æ SO

2

catalytic

converter is outside its

optimal operating

range.

Bad converter heater

Bad converter temperature sensor

Bad relay controlling the converter heater

Entire relay board is malfunctioning

I

2

C buss malfunction

DARK CAL

WARNING

The Dark Cal signal is
higher than 1 000 mV.

Light leak in reaction cell

Shutter solenoid is not functioning
Failed relay board

I

2

C bus failure

Loose connector/wiring

PMT preamp board bad or out of cal

DATA

INITIALIZED

Data Storage in iDAS

was erased

Failed disk on chip

User cleared data

FRONT PANEL

WARN

The CPU is unable to

Communicate with the
Front Panel Display

/Keyboard

Warning only appears on serial I/O com port(s)

Front panel display will be frozen, blank or will not respond.
Failed keyboard

I

2

C buss failure

Loose connector/wiring

HVPS WARNING

High voltage power

supply output is <400

V or >900 V

High voltage power supply is bad

High voltage power supply is out of cal

A/D converter circuitry is bad

IZS TEMP

WARNING

On units with IZS

options installed: The

permeation tube

temperature is Sample

chamber temperature

is

< 45°C or > 55°C

Bad IZS heater

Bad IZS temperature sensor

Bad relay controlling the IZS heater

Entire relay board is malfunctioning

I

2

C buss malfunction

Failure of thermistor interface circuitry on motherboard

PMT DET

WARNING

PMT detector output is
> 4995 mV

Failed PMT

Malfunctioning PMR preamp board

A/D converter circuitry failure

PMT TEMP
WARNING

PMT temperature is

Sample chamber

temperature is

< 2°C or > 12°C

Bad PMT thermo-electric cooler

Failed PMT TEC driver circuit

Bad PMT preamp board

Failed PMT temperature sensor

Loose wiring between PMT temperature sensor and PMT Preamp board

Malfunction of analog sensor input circuitry on motherboard

RCELL TEMP

WARNING

Sample chamber

temperature is

< 45°C or > 55°C

Bad reaction cell heater

Bad reaction cell temperature sensor

Bad relay controlling the reaction cell heater

Entire relay board is malfunctioning

I

2

C buss malfunction