beautypg.com

Using the diagnostic signal i/o function – Teledyne 6200E - Sulfides Analyzer User Manual

Page 243

background image

Model 6200E Instruction Manual

TROUBLESHOOTING & REPAIR

M6200E Rev: A1

243

Table 11-2: Test Functions - Possible Causes for Out-Of-Range Values

TEST FUNCTION

INDICATED FAILURE(S)

H2S STB

1

Unstable concentrations; leaks

SAMPLE FL

Leaks; clogged critical flow orifice

PMT

Calibration error; HVPS problem; PMT problem; No flow (leaks)

NORM PMT

Calibration error; HVPS problem; PMT problem

AZERO

Leaks; malfunctioning AutoZero valve

HVPS

HVPS broken; preamp board circuit problems

RCELL TEMP

Malfunctioning heater; relay board communication (I

2

C bus); relay

burnt out

BOX TEMP

Environment out of temperature operating range; broken

thermistor; runaway heater

PMT TEMP

TEC cooling circuit broken; High chassis temperature; 12V power

supply

IZS TEMP

(OPTION)

Malfunctioning heater; relay board communication (I

2

C bus); relay

burnt out

CONV TEMP

Malfunctioning heater or temperature sensor; relay board

communication (I

2

C bus); relay burnt out

PRESS (RX CELL

PRESSURE)

Leak; malfunctioning valve; malfunctioning pump; clogged flow

orifices; sample inlet overpressure;

H2S SLOPE

1

Calibration error; span gas concentration incorrect; leaks; low lamp
output

H2S OFFS

1

Incorrect span gas concentration/contaminated zero air/leak; low-

level calibration off

TIME OF DAY

Internal clock drifting; move across time zones; daylight savings

time?

1

Shown as they appear when analyzer is in H

2

S mode. In SO

2

mode appear as SO2 STB,

SO2 OFFS

& SO2 SLOPE. In multigas mode, both versions appear.

11.1.3. Using the Diagnostic Signal I/O Function

The signal I/O parameters found under the diagnostics (DIAG) menu combined with a thorough
understanding of the instrument’s theory of operation (Section 10) are useful for troubleshooting
in three ways:

• The technician can view the raw, unprocessed signal level of the analyzer’s critical inputs

and outputs.

• All of the components and functions that are normally under instrument control can be

manually changed.

• Analog and digital output signals can be manually controlled.

This allows the user to systematically observe the effect of these functions on the operation of
the analyzer.

Figure 11-2 shows an example of how to use the signal I/O menu to view the raw voltage of an

input signal or to control the state of an output voltage or control signal. The specific parameter
will vary depending on the situation. Please note that the analyzer will freeze it’s concentration