Figure 6-5: setup for calibrating analog outputs – Teledyne 6200E - Sulfides Analyzer User Manual
Page 90
Operating Instructions
Model 6200E Instruction Manual
90
M6200E Rev: A1
V
+DC Gnd
V OUT +
V OUT -
V IN +
V IN -
Recording
Device
ANALYZER
See Table 3-1 for
pin assignments
of Analog Out
connector on the
rear panel
Figure 6-5: Setup for Calibrating Analog Outputs
To make these adjustments, the AOUT auto-calibration feature must be turned off (Section
6.9.4.2). Activate the ANALOG I/O CONFIGURATION MENU (see Section 6.9.1), then press:
Press SET> to select the analog output channel to be configured:
DISPLAYED
AS
=
CHANNEL
CONC_OUT_1
=
A1
CONC_OUT_2
=
A2
TEST
OUTPUT
=
A4
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
< SET
SET>
CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO
CONC_OUT_1 :5V, NO CAL
< SET SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_1 CALIBRATED: YES
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_1 RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_1 VOLT–Z : 0 mV
U100 UP10 UP DOWN DN10 D100
ENTR EXIT
These keys increase / decrease the analog
output by 100, 10 or 1 counts.
Continue adjustments until the voltage measured
at the output of the analyzer and/or the input of
the recording device matches the value in the
upper right hand corner of the display to the
tolerance listed in Table 6-10.
The concentration display will not change. Only
the voltage reading of your voltmeter will change.
DIAG AIO
CONC_OUT_1 VOLT–S : 4500 mV
U100 UP10 UP DOWN DN10 D100
ENTR EXIT
DIAG AIO
CONC_OUT_1 REC OFS: 0 mV
< SET
SET>
EDIT
EXIT
If AutoCal is ON, go to
Section 6.7.3
EXIT ignores the
new setting.
ENTR accepts the
new setting.
DIAG AIO
CONC_OUT_1 AUTO CAL: OFF
< SET
SET>
EDIT
EXIT