Getting Started

Model 6200E Instruction Manual

28

M6200E Rev: A1

MODEL

6200E

Sample

Zero Air

Exhaust

Span

MODEL 701

Zero Air

Generator

VENT

Calibrated H

2

S GAS

(At high concentration)

Source of

SAMPLE Gas

Removed

during

Calibration

MODEL 700 Gas

Dilution

Calibrator

(with Ozone Bench

Option)

Figure 3-6: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator

Valve

Needle valve to control flow

MODEL

6200E

Sample

Zero Air

Exhaust

Span

MODEL 701

Zero Air

Generator

VENT

Calibrated SO

2

or H

2

S GAS

(At span gas concentration)

Source of

SAMPLE Gas

Removed

during

calibration

Figure 3-7: Pneumatic Connections–Basic Configuration–Using Bottled Span Gas

1. Attach the 1/4" exhaust line to the exhaust port of the analyzer and to the inlet port of the

pump.

CAUTION

The exhaust from the external pump needs to be vented outside the

immediate area or shelter surrounding the instrument and conform to all

safety requirements using a maximum of 10 meters of 1/4” PTFE tubing.

2. Attach the sample line to the sample inlet port. Ideally, the pressure of the sample gas should

be equal to ambient atmospheric pressure.