3.3.2. Interferences for H

2

S Measurements

It should be noted that the fluorescence method for detecting H

2

S is subject to interference from a

number of sources. Since the M6200E converts H

2

S into SO

2

and measures the UV fluorescence of

the SO

2

, the most common source of interference is from other gases that fluoresce in a similar

fashion to SO

2

, when exposed to UV Light such poly-nuclear aromatics (PNA), of which certain

hydrocarbons such as meta-xylene and naphthalene are the most pervasive. The M6200E has

been successfully tested for its ability to reject interference from most of these sources.

For a more detailed discussion of this topic, see Section 10.2.6.