Editing idas data channels – Teledyne 6200E - Sulfides Analyzer User Manual
Page 126
Operating Instructions
Model 6200E Instruction Manual
126
M6200E Rev: A1
6.11.2.2. Editing iDAS Data Channels
IDAS configuration is most conveniently done through the APICOM remote control program. The
following list of key strokes shows how to edit using the front panel.
Main Data Acquisition Menu
Edit Data Channel Menu
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
0) CONC: ATIMER, 2, 4032, R
PREV NEXT
INS DEL EDIT PRNT EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG
DAS
RNGE
PASS CLK MORE
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
EXIT will return to the
previous SAMPLE
display.
SETUP X.X
DATA ACQUISITION
VIEW
EDIT
EXIT
Moves the
display up &
down the list of
Data Channels
Inserts a new Data
Channel into the list
BEFORE the Channel
currently being displayed
Deletes The Data
Channel currently
being displayed
Exports the
configuration of all
data channels to
RS-232 interface.
Exits to the Main
Data Acquisition
Menu
SETUP X.X
NAME:CONC
EDIT PRNT
EXIT
Moves the display
between the
PROPERTIES for this
data channel.
Reports the configuration of current
data channels to the RS-232 ports.
Exits returns to the
previous Menu
Allows to edit the channel name, see next key sequence.
When editing the data channels, the top line of the display indicates some of the configuration
parameters. For example, the display line:
0) CONC : ATIMER, 4, 800
translates to the following configuration:
Channel No.: 0
NAME: CONC