beautypg.com

Manual calibration with izs option – Teledyne 6200E - Sulfides Analyzer User Manual

Page 167

background image

Model 6200E Instruction Manual

Calibration Procedures

M6200E Rev: A1

167

7.5. Manual Calibration with IZS Option

The following section describes the basic method for manually calibrating the Model 6200E analyzer
in H

2

S measurement mode using the IZS option. The same method may be used to calibrate the

M6200E analyzers configured for SO

2

measurement but requires that an SO

2

permeation tube be

installed in the instrument instead of the standard H

2

S tube.

See Section 7.8 for instructions on calibrating analyzers configured for multigas measurement
mode.

Under the best conditions, the accuracy off the H

2

S effusion rate of the IZS option’s permeation

tube is about ±5%. This can be subject to significant amounts of drift as the tube ages and the
amount of H

2

S contained in the tube is depleted. Whereas this may be sufficient for calibrating

instrument configured for H

2

S measurement and for informal calibration checks of instruments

measuring SO

2

, it is not adequate for formal SO

2

calibrations and is not approved for use by the US

EPA as a calibration source for calibrating SO

2

monitoring equipment.

For applications where more stringent calibration requirements are specified for an instrument with

an IZS option installed the following provisions must be followed.

1. Zero air and span gas must be supplied to the analyzer through the sample gas inlet as

depicted in Figure 7-1 of Section 7.2.

2. The calibration procedure must be initiated using the CAL key, not the CALZ and CALS keys,

using the procedure defined in Section 7.2.

3. Using the CAL key does not activate the zero/span or sample/cal valves of the IZS option, thus

allowing the introduction of zero air and sample gas through the sample port from more
accurate, external sources such as a calibrated bottle of H

2

S and SO

2

or a Model 700 Dilution

Calibrator.

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X


< TST TST >

CAL CALZ CALS

SETUP

Use for formal

calibration

operations.

Use only for

informal calibration

checks.