Teledyne 6200E - Sulfides Analyzer User Manual
Page 175
Model 6200E Instruction Manual
Calibration Procedures
M6200E Rev: A1
175
Toggle keys
to set
number of
days
between
procedures
(1-367)
Toggle keys
to set
delay time for
each iteration
of the
sequence:
HH:MM
(0 – 24:00)
SETUP X.X
DELTA D AYS: 1
EDIT
EXIT
SETUP X.X
DELTA DAYS: 1
0 0 2
ENTR
EXIT
SETUP X.X
DELTA DAYS:2
SET> EDIT EXIT SETUP X.X DELTA TIM E00:00 EDIT EXIT SETUP X.X DELTA TIME: 00:00 0 0 :3 0 ENTR EXIT SETUP X.X DELTA TIEM:00:30 SET> EDIT EXIT SETUP X.X STARTING TIM E:14:15 SET> EDIT EXIT EXIT returns to the SETUP Menu SETUP X.X DURATION:15.0 M INUTES EDIT EXIT SETUP X.X DURATION 15.0MINUTES 3 0 .0 ENTR EXIT SETUP X.X DURATION:30.0 M INUTES SET> EDIT EXIT Toggle keys to set duration for each iteration of the sequence: Set in Decimal from 0.1 – 60.0 SETUP X.X CALIBRATE: OFF EDIT EXIT SETUP X.X C ALIBRAT E: OFF ON ENTR EXIT SETUP X.X CALIBRATE: ON EXIT Toggle key between Off and ON SETUP X.X SEQ 2) ZERO–SPAN, 2:00:30 MODE SET EXIT Sequence Delta Time MODE Delta Days SAMPLE ENTER SETUP PASS : 818 8 1 8 ENTR EXIT SETUP X.X PRIMARY SETUP MENU AC AL DAS RNGE PASS CLK MORE EXIT SETUP X.X SEQ 1) DISABLED NEXT MODE EXIT SETUP X.X SEQ 2) DISABLED M ODE EXIT SETUP X.X M ODE: DISABLED NEXT ENTR EXIT SETUP X.X MODE: ZERO NEXT ENTR EXIT SETUP X.X M ODE: ZERO–SPAN NEXT ENTR EXIT SETUP X.X SEQ 2) ZERO–SPAN, 1:00:00 MODE SET EXIT SETUP X.X TIM ER ENABLE: ON SET> EDIT EXIT SETUP X.X STARTING DATE: 01–JAN–02 EDIT EXIT SETUP X.X STARTING DATE: 01–JAN–02 0 4 SEP 0 3 ENTR EXIT Toggle keys to set day, m onth & year: Form at : DD-MON-Y Y SETUP X.X STARTING DATE: 04–SEP–03 SET> EDIT EXIT Default value is ON SETUP X.X STARTING DATE: 04–SEP–03 SET> EDIT EXIT SETUP X.X STARTING TIM E:00:00 EDIT EXIT SETUP X.X STARTING TIM E:00:00 1 4 : 1 5 ENTR EXIT/ Toggle keys to set time: Format : HH:MM This is a 24 hr clock . PM hours are 13 – 24. Example 2:15 PM = 14:15 SAMPLE RANGE = 500.0 PPB H2S =XXX.X SETUP With dynamic calibration turned on, the state of the internal setup variables DYN_SPAN and DYN_ZERO is set to ON and the instrument will reset the slope and offset values for the H 2 S response each time the AutoCal program runs. This continuous re-adjustment of calibration dynamic calibration is enabled, the analyzer’s test functions, slope and offset values be checked
minutes
PREV NEXT
CFG
PREV NEXT
PREV
PREV
PREV NEXT
< TST TST > CAL CALZ CALS
parameters can often mask subtle fault conditions in the analyzer. It is recommended that, if
frequently to assure high quality and accurate data from the instrument.