Model 6200E Instruction Manual

Calibration Procedures

M6200E Rev: A1

165

Step Two: Set the expected H

2

S

span gas

value:

This sequence causes the

analyzer to prompt for the

expected H

2

S span

concentration.

M-P CAL

H2S SPAN CONC: 450.0 Conc

0 0 0

4 5 .0

ENTR

EXIT

M-P CAL RANGE = 500.000 PPB

H2S

=XXX.X

< TST TST > ZERO

CONC

EXIT

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST TST >

CAL

SETUP

The H

2

S span concentration

values automatically default

to 450.0 Conc.

To change this value to the

actual concentration of the

span gas, enter the number

by pressing the key under

each digit until the expected

value appears.

The span gas concentration

should always be 90% of the

selected reporting range

EXAMPLE

Reporting range = 800 ppb

Span gas conc.= 720 ppb

EXIT ignores the new setting

and returns to the previous

display.