APPENDIX A-4: M6200E Signal I/O Definitions, Revision A.1

M6200E Rev A.1

293

SIGNAL NAME

BIT OR CHANNEL

NUMBER

DESCRIPTION

ST_ZERO_CAL

3

0 = in zero calibration
1 = not in zero

ST_SPAN_CAL

4

0 = in span calibration
1 = not in span

ST_DIAG_MODE

5

0 = in diagnostic mode
1 = not in diagnostic mode

ST_LOW_SPAN_CAL

2,3,6,7

6

0 = in low span calibration
1 = not in low span

7

Spare

B status outputs, U27, J1018, pins 1–8 = bits 0–7, default I/O address 324 hex

ST_LAMP_ALARM

0

0 = lamp intensity low
1 = lamp intensity OK

ST_DARK_CAL_ALARM

1

0 = dark cal. warning
1 = dark cal. OK

ST_FLOW_ALARM

2

0 = any flow alarm
1 = all flows OK

ST_PRESS_ALARM

3

0 = any pressure alarm
1 = all pressures OK

ST_TEMP_ALARM

4

0 = any temperature alarm
1 = all temperatures OK

ST_HVPS_ALARM

5

0 = HVPS alarm
1 = HVPS OK

6–7 Spare

Front panel I

2

C keyboard, default I

2

C address 4E hex

MAINT_MODE

5 (input)

0 = maintenance mode
1 = normal mode

LANG2_SELECT

6 (input)

0 = select second language
1 = select first language (English)

SAMPLE_LED

8 (output)

0 = sample LED on
1 = off

CAL_LED

9 (output)

0 = cal. LED on
1 = off

FAULT_LED

10 (output)

0 = fault LED on
1 = off

AUDIBLE_BEEPER

14 (output)

0 = beeper on (for diagnostic testing only)
1 = off

Relay board digital output (PCF8575), default I

2

C address 44 hex

RELAY_WATCHDOG

0

Alternate between 0 and 1 at least every 5 seconds
to keep relay board active

RCELL_HEATER

1

0 = sample chamber heater on
1 = off