Model 6200E Instruction Manual
APPENDIX A-4: M6200E Signal I/O Definitions, Revision A.1
M6200E Rev A.1
293
SIGNAL NAME
BIT OR CHANNEL
NUMBER
DESCRIPTION
ST_ZERO_CAL
3
0 = in zero calibration
1 = not in zero
ST_SPAN_CAL
4
0 = in span calibration
1 = not in span
ST_DIAG_MODE
5
0 = in diagnostic mode
1 = not in diagnostic mode
ST_LOW_SPAN_CAL
2,3,6,7
6
0 = in low span calibration
1 = not in low span
7
Spare
B status outputs, U27, J1018, pins 1–8 = bits 0–7, default I/O address 324 hex
ST_LAMP_ALARM
0
0 = lamp intensity low
1 = lamp intensity OK
ST_DARK_CAL_ALARM
1
0 = dark cal. warning
1 = dark cal. OK
ST_FLOW_ALARM
2
0 = any flow alarm
1 = all flows OK
ST_PRESS_ALARM
3
0 = any pressure alarm
1 = all pressures OK
ST_TEMP_ALARM
4
0 = any temperature alarm
1 = all temperatures OK
ST_HVPS_ALARM
5
0 = HVPS alarm
1 = HVPS OK
6–7 Spare
Front panel I
2
C keyboard, default I
2
C address 4E hex
MAINT_MODE
5 (input)
0 = maintenance mode
1 = normal mode
LANG2_SELECT
6 (input)
0 = select second language
1 = select first language (English)
SAMPLE_LED
8 (output)
0 = sample LED on
1 = off
CAL_LED
9 (output)
0 = cal. LED on
1 = off
FAULT_LED
10 (output)
0 = fault LED on
1 = off
AUDIBLE_BEEPER
14 (output)
0 = beeper on (for diagnostic testing only)
1 = off
Relay board digital output (PCF8575), default I
2
C address 44 hex
RELAY_WATCHDOG
0
Alternate between 0 and 1 at least every 5 seconds
to keep relay board active
RCELL_HEATER
1
0 = sample chamber heater on
1 = off