Model 6200E Instruction Manual

APPENDIX A-3: Warnings and Test Functions, Revision A.1

M6200E Rev A.1

291

Table A-3: M6200E Test Functions, Revision A.1

TEST Function

Message Text

DESCRIPTION

RANGE

RANGE=500.0 PPB

3

D/A range in single or auto-range modes.

RANGE1

RANGE1=500.0 PPB

3

D/A #1 range in independent range mode.

RANGE2

RANGE2=500.0 PPB

3

D/A #2 range in independent range mode.

STABILITY

H2S STB

1

=0.0 PPB

3, 4

Concentration stability (standard deviation based on
setting of STABIL_FREQ and STABIL_SAMPLES).

RESPONSE

2

RSP=1.11(0.00) SEC

Instrument response. Length of each signal processing
loop. Time in parenthesis is standard deviation.

VACUUM

5

VAC=9.1 IN-HG-A

Vacuum pressure.

SAMPPRESS

PRES=29.9 IN-HG-A

Sample pressure.

SAMPFLOW

SAMP FL=700 CC/M

Sample flow rate.

PMTDET

PMT=762.5 MV

Raw PMT reading.

NORMPMTDET

NORM PMT=742.9 MV

PMT reading normalized for temperature, pressure,
auto-zero offset, but not range.

UVDET

UV LAMP=3457.6 MV

UV lamp reading.

LAMPRATIO

LAMP RATIO=100.0 % UV lamp ratio of current reading divided by calibrated

reading.

STRAYLIGHT

STR. LGT=0.1 PPB

Stray light offset.

DARKPMT

DRK PMT=19.6 MV

PMT dark offset.

DARKLAMP

DRK LMP=42.4 MV

UV lamp dark offset.

SLOPE

H2S SLOPE

4

=1.061

Slope for current range, computed during zero/span

calibration.

OFFSET

H2S OFFS

4

=250.0 MV

Offset for current range, computed during zero/span
calibration.

HVPS

HVPS=650 VOLTS

High voltage power supply output.

RCELLDUTY

RCELL ON=0.00 SEC

Sample chamber temperature control duty cycle.

RCELLTEMP

RCELL TEMP=52.1 C

Sample chamber temperature.

BOXTEMP

BOX TEMP=35.5 C

Internal chassis temperature.

PMTTEMP

PMT TEMP=7.0 C

PMT temperature.

IZSDUTY

IZS ON=0.00 SEC

IZS temperature control duty cycle.

IZSTEMP

IZS TEMP=52.2 C

IZS temperature.

CONVTEMP

CONV TEMP=315.0 C

H

2

S Æ SO

2

Converter temperature.

H

2

S

H

2

S=261.4 PPB

H

2

S concentration for current range.

TESTCHAN

TEST=3721.1 MV

Value output to TEST_OUTPUT analog output, selected

with TEST_CHAN_ID variable.

CLOCKTIME

TIME=10:38:27

Current instrument time of day clock.

1

The name is used to request a message via the RS-232 interface, as in “T BOXTEMP”.

2

Engineering software.

3

Current instrument units.

4

Shown as they appear when analyzer is in H

2

S mode. In SO

2

mode appear as SO2 STB, SO2 OFFS &

SO2 SLOPE. In multigas mode, both versions appear.

5