APPENDIX A-3: Warnings and Test Functions, Revision A.1
M6200E Rev A.1
291
Table A-3: M6200E Test Functions, Revision A.1
TEST Function
Message Text
DESCRIPTION
RANGE
RANGE=500.0 PPB
3
D/A range in single or auto-range modes.
RANGE1
RANGE1=500.0 PPB
3
D/A #1 range in independent range mode.
RANGE2
RANGE2=500.0 PPB
3
D/A #2 range in independent range mode.
STABILITY
H2S STB
1
=0.0 PPB
3, 4
Concentration stability (standard deviation based on
setting of STABIL_FREQ and STABIL_SAMPLES).
RESPONSE
2
RSP=1.11(0.00) SEC
Instrument response. Length of each signal processing
loop. Time in parenthesis is standard deviation.
VACUUM
5
VAC=9.1 IN-HG-A
Vacuum pressure.
SAMPPRESS
PRES=29.9 IN-HG-A
Sample pressure.
SAMPFLOW
SAMP FL=700 CC/M
Sample flow rate.
PMTDET
PMT=762.5 MV
Raw PMT reading.
NORMPMTDET
NORM PMT=742.9 MV
PMT reading normalized for temperature, pressure,
auto-zero offset, but not range.
UVDET
UV LAMP=3457.6 MV
UV lamp reading.
LAMPRATIO
LAMP RATIO=100.0 % UV lamp ratio of current reading divided by calibrated
reading.
STRAYLIGHT
STR. LGT=0.1 PPB
Stray light offset.
DARKPMT
DRK PMT=19.6 MV
PMT dark offset.
DARKLAMP
DRK LMP=42.4 MV
UV lamp dark offset.
SLOPE
H2S SLOPE
4
=1.061
Slope for current range, computed during zero/span
calibration.
OFFSET
H2S OFFS
4
=250.0 MV
Offset for current range, computed during zero/span
calibration.
HVPS
HVPS=650 VOLTS
High voltage power supply output.
RCELLDUTY
RCELL ON=0.00 SEC
Sample chamber temperature control duty cycle.
RCELLTEMP
RCELL TEMP=52.1 C
Sample chamber temperature.
BOXTEMP
BOX TEMP=35.5 C
Internal chassis temperature.
PMTTEMP
PMT TEMP=7.0 C
PMT temperature.
IZSDUTY
IZS ON=0.00 SEC
IZS temperature control duty cycle.
IZSTEMP
IZS TEMP=52.2 C
IZS temperature.
CONVTEMP
CONV TEMP=315.0 C
H
2
S Æ SO
2
Converter temperature.
H
2
S
H
2
S=261.4 PPB
H
2
S concentration for current range.
TESTCHAN
TEST=3721.1 MV
Value output to TEST_OUTPUT analog output, selected
with TEST_CHAN_ID variable.
CLOCKTIME
TIME=10:38:27
Current instrument time of day clock.
1
The name is used to request a message via the RS-232 interface, as in “T BOXTEMP”.
2
Engineering software.
3
Current instrument units.
4
Shown as they appear when analyzer is in H
2
S mode. In SO
2
mode appear as SO2 STB, SO2 OFFS &
SO2 SLOPE. In multigas mode, both versions appear.
5
M6200EH