beautypg.com

Electrical test – Teledyne 6200E - Sulfides Analyzer User Manual

Page 96

background image

Operating Instructions

Model 6200E Instruction Manual

96

M6200E Rev: A1

6.9.6. Electrical Test

The electrical test function creates a current, which substitutes the PMT signal, and feeds it into the

preamplifier board. This signal is generated by circuitry on the pre-amplifier board itself and tests
the filtering and amplification functions of that assembly along with the A/D converter on the

motherboard. It does not test the PMT itself. The electrical test should produce a PMT signal of
about 2000 ±1000 mV.

To activate the electrical test press the following keys.

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

DIAG

SIGNAL I / O

PREV

NEXT

JUMP

ENTR

EXIT

DIAG ELEC

RANGE = 500.000 PPB O2=X.XXX

EXIT

Press NEXT until…

DIAG

ELECTRICAL TEST

PREV NEXT

ENTR

EXIT

Press TST until…

DIAG ELEC

PMT = 1732 MV

SO2=X.XXX

EXIT

While the electrical test is

activated, PMT should equal:

2000 mV ± 1000 mV