APPENDIX A-2: Setup Variables For Serial I/O, Revision A.1
Model 6200E Instruction Manual
288
M6200E Rev A.1
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
cycle period.
RCELL_PROP
1/ºC 0.3
(prop.
band =
3.3 ºC)
0–10 Reaction
cell
temperature
PID
proportional coefficient.
RCELL_INTEG
— 0.005 0–10
Reaction
cell
temperature
PID
integral coefficient.
RCELL_DERIV
— 0.5
0–10
Reaction
cell
temperature
PID
derivative coefficient.
IZS_CYCLE
Seconds 2
0.5–30
IZS temperature control cycle period.
IZS_PROP
1/ºC 1
(prop.
band = 1
ºC)
0–10
IZS temperature PID proportional
coefficient.
IZS_INTEG
—
0.03
0–10
IZS temperature PID integral
coefficient.
IZS_DERIV
—
0
0–10
IZS temperature PID derivative
coefficient.
HVPS_SET
Volts 650
Warnings
:
400–900
0–2000
High voltage power supply warning
limits. Set point is not used.
DETECTOR_LIMIT
mV 1000
Warnings
:
600–
4995
0–5000
UV lamp and PMT detector warning
limits. Set point is not used.
SERIAL_NUMBER
—
“00000
000 ”
Any character in
the allowed
character set. Up
to 100 characters
long.
Unique serial number for instrument.
Enclose value in double quotes (“)
when setting from the RS-232
interface.
DISP_INTENSITY
— HIGH HIGH,
MED,
LOW,
DIM
Front panel display intensity. Enclose
value in double quotes (“) when
setting from the RS-232 interface.
I2C_RESET_ENABLE
— ON
OFF,
ON
I
2
C bus automatic reset enable.