APPENDIX A-2: Setup Variables For Serial I/O, Revision A.1

Model 6200E Instruction Manual

288

M6200E Rev A.1

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT

VALUE

VALUE RANGE

DESCRIPTION

cycle period.

RCELL_PROP

1/ºC 0.3

(prop.

band =
3.3 ºC)

0–10 Reaction

cell

temperature

PID

proportional coefficient.

RCELL_INTEG

— 0.005 0–10

Reaction

cell

temperature

PID

integral coefficient.

RCELL_DERIV

— 0.5

0–10

Reaction

cell

temperature

PID

derivative coefficient.

IZS_CYCLE

Seconds 2

0.5–30

IZS temperature control cycle period.

IZS_PROP

1/ºC 1

(prop.
band = 1

ºC)

0–10

IZS temperature PID proportional

coefficient.

IZS_INTEG

0.03

0–10

IZS temperature PID integral
coefficient.

IZS_DERIV

0

0–10

IZS temperature PID derivative
coefficient.

HVPS_SET

Volts 650

Warnings

:
400–900

0–2000

High voltage power supply warning
limits. Set point is not used.

DETECTOR_LIMIT

mV 1000

Warnings
:
600–
4995

0–5000

UV lamp and PMT detector warning
limits. Set point is not used.

SERIAL_NUMBER

“00000
000 ”

Any character in
the allowed

character set. Up
to 100 characters

long.

Unique serial number for instrument.
Enclose value in double quotes (“)

when setting from the RS-232
interface.

DISP_INTENSITY

— HIGH HIGH,

MED,
LOW,
DIM

Front panel display intensity. Enclose
value in double quotes (“) when
setting from the RS-232 interface.

I2C_RESET_ENABLE

— ON

OFF,

ON

I

2

C bus automatic reset enable.