Model 6200E Instruction Manual

APPENDIX A-2: Setup Variables For Serial I/O, Revision A.1

M6200E Rev A.1

287

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT

VALUE

VALUE RANGE

DESCRIPTION

&B0 &N6
&M0 E0

Q1 &W0”

long.

quotes (“) when setting from the RS-
232 interface.

RS232_PASS

Password 940331

0–999999

RS-232 log on password.

MACHINE_ID

ID 101

2

,

102

3

,

108

4

0–9999

Unique ID number for instrument.

COMMAND_PROMPT

“Cmd> ”

Any character in

the allowed
character set. Up

to 100 characters

long.

RS-232 interface command prompt.

Displayed only if enabled with
RS232_MODE variable. Enclose value

in double quotes (“) when setting

from the RS-232 interface.

TEST_CHAN_ID

— NONE

NONE,
PMT READING,
UV READING,
SAMPLE
PRESSURE,
SAMPLE

FLOW,
RCELL TEMP,
CHASSIS
TEMP,
IZS TEMP,
PMT TEMP

,

CONV TEMP

2

,

HVPS

VOLTAGE

Diagnostic analog output ID. Enclose

value in double quotes (“) when
setting from the RS-232 interface.

REMOTE_CAL_MODE

— SO2-

LOW

SO2-LOW,
SO2-HIGH,
H2S-LOW,
H2S-HIGH

Gas and range to calibrate during
contact-closure and Hessen

calibration. Enclose value in double
quotes (“) when setting from the RS-

232 interface.

HOLD_DAC_ON_CAL

OFF

OFF, ON

ON holds D/A outputs during
zero/span calibration; OFF permits

D/A outputs to change.

PASS_ENABLE

OFF

OFF, ON

ON enables passwords; OFF disables
them.

STABIL_GAS

— SO2 SO2,

H2S

Gas to use to measure concentration
stability. Enclose value in double
quotes (") when setting from the RS-

232 interface.

STABIL_FREQ

Seconds 10

1–300

Stability

measurement

sampling

period.

STABIL_SAMPLES

Samples 25

2–40

Number of samples in concentration

stability reading.

RCELL_CYCLE

Seconds 2

0.5–30

Reaction

cell

temperature

control