Model 6200E Instruction Manual

Calibration Procedures

M6200E Rev: A1

169

STEP ONE: Connect the sources of Zero Air and Span Gas as shown below.


External Zero

Air Scrubber

Filter

MODEL 701

Zero Air

Generator

Source of

SAMPLE Gas

VENT if input is pressurized

Zero/Span Valves – Option 50

MODEL 700

Gas Dilution Calibrator

(with O

3

generator option)

VENT

MODEL

6200E

Sample

Zero Air

Exhaust

Span

Source of

SAMPLE Gas

VENT if input is pressurized

Internal Zero/Span Option (IZS) – Option 51


MODEL

6200E

Sample

Zero Air

Exhaust

Span

Calibrated H

2

S gas

(At high concentration)

Ambient

Air

Figure 7-3: Setup for Manual Calibration Check with Z/S Valve or IZS Option