6.12.4.6. Hessen Protocol Gas ID

Since the M6200E can be, when the proper optional equipment is installed and operating, a
multigas instrument that measures both H

2

S and SO

2

, both of these gases are listed in the Hessen

protocol gas list. In its default state the Hessen protocol firmware assigns both gases a Hessen ID

number and actively reports both even if the instrument is only measuring one.

To change or edit these settings press:

KEY FUNCTION

Moves to next gas entry in list

NEXT>

Moves the cursor previous gas entry in list

INS

Inserts a new gas entry into the list.

DEL

Deletes the >>>>>.

ENTR

Accepts the new setting and returns to the previous menu.

EXIT

Ignores the new setting and returns to the previous menu.

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS DIAG ALRM

EXIT

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST TST > CAL

SETUP

SETUP X.X

HESSEN RESPONSE MODE :CMD

SET>

EDIT

EXIT

SETUP X.X

COMMUNICATIONS MENU

ID

HESN

COM1

COM2

EXIT

SETUP X.X

HESSEN VARIATION: TYPE 1

SET>

EDIT

EXIT

SETUP X.X

HESSEN GAS LIST

SET>

EDIT

EXIT

ENTR key

accepts the new

settings

EXIT key ignores

the new settings

SETUP X.X

SO2, 111, REPORTED

NEXT>

INS

DEL

EDIT

PRNT EXIT

SETUP X.X

GAS TYPE SO2

NEXT>

ENTR

EXIT

Use the PREV & NEXT keys to cycle

between the SO2 & H2S

SETUP X.X

GAS ID: 111

0 0 0

ENTR

EXIT

Use the PREV & NEXT keys to cycle

between the SO2 & H2S

SETUP X.X

REPORTED : ON

ON

ENTR

EXIT

Toggle this key to switch reporting

Between ON and OFF

Use the PREV & NEXT keys to cycle

existing entries in Hessen gas list