6.12.4.6. Hessen Protocol Gas ID
Since the M6200E can be, when the proper optional equipment is installed and operating, a
multigas instrument that measures both H
2
S and SO
2
, both of these gases are listed in the Hessen
protocol gas list. In its default state the Hessen protocol firmware assigns both gases a Hessen ID
number and actively reports both even if the instrument is only measuring one.
To change or edit these settings press:
KEY FUNCTION
Moves to next gas entry in list
NEXT>
Moves the cursor previous gas entry in list
INS
Inserts a new gas entry into the list.
DEL
Deletes the >>>>>.
ENTR
Accepts the new setting and returns to the previous menu.
EXIT
Ignores the new setting and returns to the previous menu.
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG ALRM
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
HESSEN RESPONSE MODE :CMD
SET> EDIT EXIT SETUP X.X COMMUNICATIONS MENU ID HESN COM1 COM2 EXIT SETUP X.X HESSEN VARIATION: TYPE 1 SET> EDIT EXIT SETUP X.X HESSEN GAS LIST SET> EDIT EXIT ENTR key accepts the new settings EXIT key ignores the new settings SETUP X.X SO2, 111, REPORTED NEXT> INS DEL EDIT PRNT EXIT SETUP X.X GAS TYPE SO2 NEXT> ENTR EXIT Use the PREV & NEXT keys to cycle between the SO2 & H2S SETUP X.X GAS ID: 111 0 0 0 ENTR EXIT Use the PREV & NEXT keys to cycle between the SO2 & H2S SETUP X.X REPORTED : ON ON ENTR EXIT Toggle this key to switch reporting Between ON and OFF Use the PREV & NEXT keys to cycle existing entries in Hessen gas list