background image

Getting Started

Model 6200E Instruction Manual

26

M6200E Rev: A1

3.1.1.3. Connecting the Control Inputs

If you wish to use the analyzer to remotely activate the zero and span calibration modes, several
digital control inputs are provided through a 10-pin connector labeled CONTROL IN on the

analyzer’s rear panel.

There are two methods for energizing the control inputs. The internal +5V available from the pin

labeled “+” is the most convenient method. However, if full isolation is required, an external 5
VDC power supply should be used.

SP

AN

C

AL

ZE

R

O

CA

L

SP

AN

C

AL

ZE

R

O

CA

L

CONTROL IN

A B C D E F U

+

CONTROL IN

A B C D E F U

+

-

+

Local Power Connections

External Power Connections

5 VDC Power

Supply

Figure 3-5: Control Input Connector

Table 3-3: Control Input Signals

INPUT #

STATUS

DEFINITION

ON CONDITION

A

REMOTE ZERO CAL

The analyzer is placed in Zero Calibration mode. The mode

field of the display will read ZERO CAL R.

B

REMOTE

LO SPAN CAL

The analyzer is placed in low span calibration mode as part
of performing a low span (midpoint) calibration. The mode

field of the display will read LO CAL R.

C, D, E & F

SPARE

Digital Ground

The ground level from the analyzer’s internal DC power
supplies (same as chassis ground)

U

External Power

input

Input pin for +5 VDC required to activate pins A – F.

+

5 VDC output

Internally generated 5V DC power. To activate inputs A – F,
place a jumper between this pin and the “U” pin. The

maximum amperage through this port is 300 mA (combined
with the analog output supply, if used).