3.1.1.3. Connecting the Control Inputs
If you wish to use the analyzer to remotely activate the zero and span calibration modes, several
digital control inputs are provided through a 10-pin connector labeled CONTROL IN on the
analyzer’s rear panel.
There are two methods for energizing the control inputs. The internal +5V available from the pin
labeled “+” is the most convenient method. However, if full isolation is required, an external 5
VDC power supply should be used.
SP
AN
C
AL
ZE
R
O
CA
L
SP
AN
C
AL
ZE
R
O
CA
L
CONTROL IN
A B C D E F U
+
CONTROL IN
A B C D E F U
+
-
+
Local Power Connections
External Power Connections
5 VDC Power
Supply
Figure 3-5: Control Input Connector
Table 3-3: Control Input Signals
INPUT #
STATUS
DEFINITION
ON CONDITION
A
REMOTE ZERO CAL
The analyzer is placed in Zero Calibration mode. The mode
field of the display will read ZERO CAL R.
B
REMOTE
LO SPAN CAL
The analyzer is placed in low span calibration mode as part
of performing a low span (midpoint) calibration. The mode
field of the display will read LO CAL R.
C, D, E & F
SPARE
Digital Ground
The ground level from the analyzer’s internal DC power
supplies (same as chassis ground)
U
External Power
input
Input pin for +5 VDC required to activate pins A – F.
+
5 VDC output
Internally generated 5V DC power. To activate inputs A – F,
place a jumper between this pin and the “U” pin. The
maximum amperage through this port is 300 mA (combined
with the analog output supply, if used).