Teledyne 6200E - Sulfides Analyzer User Manual
Page 290
APPENDIX A-3: Warnings and Test Functions, Revision A.1
Model 6200E Instruction Manual
290
M6200E Rev A.1
APPENDIX A-3: Warnings and Test Functions, Revision A.1
Table A-2: M6200E Warning Messages, Revision A.1
NAME MESSAGE
TEXT
DESCRIPTION
WSYSRES SYSTEM
RESET
Instrument was power-cycled or the CPU was reset.
WDATAINIT DATA
INITIALIZED
Data storage was erased.
WCONFIGINIT CONFIG
INITIALIZED
Configuration storage was reset to factory configuration
or erased.
WPMT
PMT DET WARNING
PMT detector outside of warning limits specified by
DETECTOR_LIMIT variable.
WUVLAMP
UV LAMP WARNING
UV lamp reading outside of warning limits specified by
DETECTOR_LIMIT variable.
WSAMPFLOW SAMPLE
FLOW
WARN
Sample flow outside of warning limits specified by
SAMP_FLOW_SET variable.
WSAMPPRESS
SAMPLE PRESS WARN
Sample pressure outside of warning limits specified by
SAMP_PRESS_SET variable.
WVACPRESS
VACUUM PRESS WARN Vacuum pressure outside of warning limits specified by
VAC_PRESS_SET variable.
WCONVTEMP
CONV TEMP WARNING
Converter temperature outside of warning limits specified
by CONV_SET variable.
WBOXTEMP BOX
TEMP
WARNING
Chassis temperature outside of warning limits specified
by BOX_SET variable.
WRCELLTEMP RCELL
TEMP
WARNING
Sample chamber temperature outside of warning limits
specified by RCELL_SET variable.
WIZSTEMP
IZS TEMP WARNING
IZS temperature outside of warning limits specified by
IZS_SET variable.
WPMTTEMP PMT
TEMP
WARNING
PMT temperature outside of warning limits specified by
PMT_SET variable.
WDARKCAL
DARK CAL WARNING
Dark offset above limit specified by DARK_LIMIT
variable.
WHVPS HVPS
WARNING
High voltage power supply output outside of warning
limits specified by HVPS_SET variable.
WDYNZERO
CANNOT DYN ZERO
Contact closure zero calibration failed while DYN_ZERO
was set to ON.
WDYNSPAN
CANNOT DYN SPAN
Contact closure span calibration failed while DYN_SPAN
was set to ON.
WREARBOARD
REAR BOARD NOT DET Rear board was not detected during power up.
WRELAYBOARD
RELAY BOARD WARN
Firmware is unable to communicate with the relay board.
WFRONTPANEL FRONT
PANEL
WARN
Firmware is unable to communicate with the front panel.
WANALOGCAL ANALOG
CAL
WARNING
The A/D or at least one D/A channel has not been
calibrated.