2.2. EPA Equivalency Designation
The Model 6200E Analyzer is designated as Reference Method Number EQOA-XXXX-XXX as per 40
CFR Part 53 when operated under the following conditions:
• Range: Any range from 50 parts per billion (ppb) to 10 parts per million (ppm).
• Ambient temperature range of 5
o
C to 40
o
C.
• Line voltage range of 105-125 VAC or 220-240 VAC, at 50 or 60 Hz.
• Sample filter: Equipped with PTFE filter element in the internal filter assembly.
• Sample flow of 650 +/- 65 cc/min.
• Vacuum pump (internal or external) capable of 14"Hg absolute pressure @ 1 slpm or
better.
• Software settings:
Dynamic span
OFF
Dynamic zero
OFF
Dilution factor
OFF
AutoCal ON
or
OFF
IND range
ON or OFF
Auto-range ON
or
OFF
Temp/Pressure compensation
ON
Under the designation, the analyzer may be operated with or without the following optional
equipment:
• Rack mount with or without chassis slides.
• Zero/span valve options.
• Internal zero/span (IZS) option with:
• H
2
S permeation tube - 0.4ppm at 0.7 liter per minute; certified/uncertified.
• H
2
S permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified. Under the
designation, the IZS option cannot be used as the source of calibration.
• 4-20mA isolated analog outputs.
• Status outputs.
• Control inputs.
• RS-232 output.
• Ethernet output.
• Zero air scrubber.
• 4-20mA, isolated output.