2.2. EPA Equivalency Designation

The Model 6200E Analyzer is designated as Reference Method Number EQOA-XXXX-XXX as per 40

CFR Part 53 when operated under the following conditions:

• Range: Any range from 50 parts per billion (ppb) to 10 parts per million (ppm).

• Ambient temperature range of 5

o

C to 40

o

C.

• Line voltage range of 105-125 VAC or 220-240 VAC, at 50 or 60 Hz.

• Sample filter: Equipped with PTFE filter element in the internal filter assembly.

• Sample flow of 650 +/- 65 cc/min.

• Vacuum pump (internal or external) capable of 14"Hg absolute pressure @ 1 slpm or

better.

• Software settings:

Dynamic span

OFF

Dynamic zero

OFF

Dilution factor

OFF

AutoCal ON

or

OFF

IND range

ON or OFF

Auto-range ON

or

OFF

Temp/Pressure compensation

ON


Under the designation, the analyzer may be operated with or without the following optional

equipment:

• Rack mount with or without chassis slides.

• Zero/span valve options.

• Internal zero/span (IZS) option with:

• H

2

S permeation tube - 0.4ppm at 0.7 liter per minute; certified/uncertified.

• H

2

S permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified. Under the

designation, the IZS option cannot be used as the source of calibration.

• 4-20mA isolated analog outputs.

• Status outputs.

• Control inputs.

• RS-232 output.

• Ethernet output.

• Zero air scrubber.

• 4-20mA, isolated output.