beautypg.com

Analyzer status led’s, Keyboard, Display – Teledyne 6200E - Sulfides Analyzer User Manual

Page 233

background image

Model 6200E Instruction Manual

Theory Of Operation

M6200E Rev: A1

233

The most commonly used method for communicating with the M6200E UV Fluorescence H

2

S

Analyzer is via the instrument’s front panel which includes a set of three status LEDs, a vacuum
florescent display and a keyboard with 8 context-sensitive keys.

10.6.1.1. Analyzer Status LED’s

Three LEDS are used to inform the user of the instruments basic operating status.

Table 10-3: Front Panel Status LED’s

NAME COLOR STATE

DEFINITION

SAMPLE Green Off

On

Blinking

Unit is not operating in sample mode, iDAS is disabled.
Sample Mode active; Front Panel Display being updated, iDAS data being stored.
Unit is operating in sample mode, front panel display being updated, iDAS hold-off

mode is ON, iDAS disabled

CAL Yellow Off

On

Blinking

Auto Cal disabled
Auto Cal enabled
Unit is in calibration mode

FAULT Red Off

Blinking

H

2

S

warnings exist

Warnings exist

10.6.1.2. Keyboard

A row of eight keys just below the vacuum florescent display (see Figure 10-20) is the main
method by which the user interacts with the analyzer. As the software is operated, labels appear
on the bottom row of the display directly above each active key, defining the function of that key
as it is relevant for the operation being performed. Pressing a key causes the associated
instruction to be performed by the analyzer.

Note that the keys do not auto-repeat. In circumstances where the same key must be activated
for two consecutive operations, it must be released and re-pressed.

10.6.1.3. Display

The main display of the analyzer is a vacuum fluorescent display with two lines of 40 text
characters each. Information is organized in the following manner (see Figure 10-20):

Mode Field: Displays the name of the analyzer’s current operating mode.

Message Field: Displays a variety of informational messages such as warning messages, operation

data and response messages during interactive tasks.

Concentration Field: Displays the actual concentration of H

2

S in the sample gas currently being

measured by the analyzer. Switches to indicate SO

2

concentration if/when the instrument is

configured for SO

2

measurement or between H

2

S and SO

2

when the instrument is in multigas

measurement mode.

Keypad Definition Field: Displays the definitions for the row of keys just below the display. These
definitions dynamic, context sensitive and software driven.