beautypg.com

Zero air, Gas standards, Permeation tubes – Teledyne 6200E - Sulfides Analyzer User Manual

Page 158

background image

Calibration Procedures

Model 6200E Instruction Manual

158

M6200E Rev: A1

7.1.2. Zero Air

Zero air is similar in chemical composition to the Earth’s atmosphere but scrubbed of all

components that might affect the analyzer’s readings. For H

2

S measuring devices, zero air should

be similar in composition to the sample gas but devoid of H

2

S, hydrocarbons, and Sulfur dioxide

(SO

2

).

Devices such as the API Model 701 zero air generator that condition ambient air by drying and

removal of pollutants are available. We recommend this type of device for generating zero air.

7.1.3. Gas Standards

Span gas is specifically mixed to match the chemical composition of the gas being measured at

about 90% of the desired full measurement range. For example, if the measurement range is 500
ppb, the span gas should have an H

2

S concentration of about 450 ppb.

We strongly recommend that span calibration is carried out with bottled, calibrated H

2

S or SO

2

span gas, although it is possible to use a permeation tube such as that included in the IZS valve
option. Span gases should be certified to a specific accuracy to ensure accurate calibration of the

analyzer. Typical gas accuracy for calibrated span gases is 1 or 2 %. H

2

S and SO

2

standard gases

should be mixed in nitrogen.

7.1.4. Permeation Tubes

Teledyne Analytical Instruments offers an IZS option operating with permeation devices. The

accuracy of these devices is about ±5%. Whereas this may be sufficient for quick, daily calibration
checks, we recommend the use of certified H

2

S gases for accurate calibration.

NOTE

Applications requiring US-EPA equivalency do not allow permeation devices to be used as

sources of span gas for calibration of the analyzer where EPA equivalency is required,

such as SO2 monitoring.