Model 6200E Instruction Manual

Operating Instructions

M6200E Rev: A1

73

6.7.4. Single Range mode (SNGL)

The default range mode for the analyzer is single range, in which all analog concentration outputs

are set to the same reporting range. This reporting range can be set to any value between 5.0 ppb
and 20 000 ppb.

While the two outputs always have the same reporting range, the span and scaling of their
electronic signals may also be configured for different differently (e.g., A1 = 0-10 V; A2 = 0-0.1

V).

To select SNGLE range mode and to set the upper limit of the range, press:



SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP C.3

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP C.3

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SETUP C.3

RANGE MODE:

SNGL

SNGL

IND AUTO

ENTR EXIT

EXIT x 2 returns

to the main

SAMPLE display

SETUP C.3

RANGE MODE:

SNGL

SNGL

IND AUTO

ENTR

EXIT

SETUP C.3

RANGE CONTROL MENU

MODE

SET

UNIT

EXIT

SETUP C.3

RANGE:

500.0 Conc

0 0

5

0 0 .0

ENTR

EXIT

SETUP C.3

RANGE CONTROL MENU

MODE

SET UNIT

EXIT

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST TST > CAL

SETUP

NOTE

On analyzers with the multigas option activated (see section 5.5 and Section 6.8.1) the

concentration value will switch back and forth between from “H2S=XXX.X” to

“SO2=XXX.X” depending on which gas is currently being measured.