beautypg.com

Teledyne 6200E - Sulfides Analyzer User Manual

Page 141

background image

Model 6200E Instruction Manual

Operating Instructions

M6200E Rev: A1

141

Table 6-23: Control Input Pin Assignments

INPUT

STATUS

CONDITION WHEN ENABLED

A

External Zero Cal Zero calibration mode is activated. The mode field of the

display will read ZERO CAL R.

B External

Span

Cal

Span calibration mode is activated. The mode field of the
display will read SPAN CAL R.

C

External Low

Span Cal

Low span (mid-point) calibration mode is activated. The mode
field of the display will read LO CAL R.

D

Unused

E

Unused

F

Unused

Digital Ground

Provided to ground an external device (e.g., recorder).

U

DC Power For

Input Pull Ups

Input for +5 VDC required to activate inputs A - F. This voltage
can be taken from an external source or from the “+” pin.

+

Internal +5v

Supply

Internal source of +5V which can be used to activate inputs
when connected to pin U.

There are two methods to activate control inputs. The internal +5V available from the “+” pin is
the most convenient method (Figure 6.18). However, to ensure that these inputs are truly isolated,

a separate, external 5 VDC power supply should be used (Figure 6.19).

CONTROL IN

A B C D E F U

+

LO

W

S

P

A

N

ZE

RO

SP

AN

Figure 6-18: Control Inputs with local 5 V power supply

CONTROL IN

-

+

5 VDC Power

Supply

LOW S

P

A

N

ZE

R

O

SP

A

N

A B C D E F U

+

Figure 6-19: Control Inputs with external 5 V power supply