Calibration Procedures

Model 6200E Instruction Manual

168

M6200E Rev: A1

7.6. Manual Calibration Checks with IZS or Zero/Span

Valves

Zero and span checks using the zero/span valve or IZS option are similar to that described in
Section 7.3, except:

On units with an IZS option installed, zero air and span gas are supplied to the analyzer through
the zero gas inlet and from ambient air.

On units with a zero/span valve option installed, zero air and span gas are supplied to the analyzer
through the zero gas and span gas inlets from two different sources.

The zero and calibration operations are initiated directly and independently with dedicated keys
CALZ and CALS.

To perform a manual calibration check of an analyzer with a zero/span valve or IZS Option

installed, use the following method:

NOTE

The instrument can only be fitted with one type of permeation tube at a time. Therefore

the IZS option can only be used to calibrate or check the instrument for one gas, H

2

S or

SO

2

, but not both.