Model 6200E Instruction Manual

APPENDIX A-2: Setup Variables For Serial I/O, Revision A.1

M6200E Rev A.1

283

APPENDIX A-2: Setup Variables For Serial I/O, Revision A.1

Table A-1: M6200E Setup Variables, Revision A.1

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT

VALUE

VALUE RANGE

DESCRIPTION

MEASURE_MODE

— SO2-H2S

1

SO2,
SO2-H2S,
H2S

Gas measurement mode. Enclose
value in double quotes (") when
setting from the RS-232 interface.

DAS_HOLD_OFF

Minutes

15

0.5–20

Duration of DAS hold off period.

TPC_ENABLE

ON

OFF, ON

ON enables temperature and
pressure compensation; OFF disables
it.

RCELL_SET

ºC 50

Warnings
:
45–55

30–70

Reaction cell temperature set point
and warning limits.

IZS_SET

ºC 50

Warnings
:
45–55

30–70

IZS temperature set point and
warning limits.

DYN_ZERO

OFF

OFF, ON

ON enables contact closure dynamic

zero; OFF disables it.

DYN_SPAN

OFF

OFF, ON

ON enables contact closure dynamic
span; OFF disables it.

CONC_PRECISION

— 1

AUTO,
0,
1,
2,
3,
4

Number of digits to display to the
right of the decimal point for

concentrations on the display.
Enclose value in double quotes (“)
when setting from the RS-232

interface.

CLOCK_ADJ

Sec./Day 0

-60–60

Time-of-day clock speed adjustment.

LANGUAGE_SELECT

— ENGL ENGL,

SECD,
EXTN

Selects the language to use for the
user interface. Enclose value in

double quotes (“) when setting from

the RS-232 interface.

MAINT_TIMEOUT

Hours

2

0.1–100

Time until automatically switching
out of software-controlled
maintenance mode.

MEASURE_PERIOD

Minutes

10

1–60

Length of time to measure each gas.

MEASURE_DELAY

Minutes

3

0.1–20

How long to defer sampling after
switching streams.

CONV_TIME

33 MS

33 MS,
66 MS,
133 MS,
266 MS,
533 MS,
1 SEC,
2 SEC

Conversion time for PMT and UV

detector channels. Enclose value in
double quotes (“) when setting from

the RS-232 interface.