Model 6200E Instruction Manual
APPENDIX A-2: Setup Variables For Serial I/O, Revision A.1
M6200E Rev A.1
283
APPENDIX A-2: Setup Variables For Serial I/O, Revision A.1
Table A-1: M6200E Setup Variables, Revision A.1
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
MEASURE_MODE
— SO2-H2S
1
SO2,
SO2-H2S,
H2S
Gas measurement mode. Enclose
value in double quotes (") when
setting from the RS-232 interface.
DAS_HOLD_OFF
Minutes
15
0.5–20
Duration of DAS hold off period.
TPC_ENABLE
—
ON
OFF, ON
ON enables temperature and
pressure compensation; OFF disables
it.
RCELL_SET
ºC 50
Warnings
:
45–55
30–70
Reaction cell temperature set point
and warning limits.
IZS_SET
ºC 50
Warnings
:
45–55
30–70
IZS temperature set point and
warning limits.
DYN_ZERO
—
OFF
OFF, ON
ON enables contact closure dynamic
zero; OFF disables it.
DYN_SPAN
—
OFF
OFF, ON
ON enables contact closure dynamic
span; OFF disables it.
CONC_PRECISION
— 1
AUTO,
0,
1,
2,
3,
4
Number of digits to display to the
right of the decimal point for
concentrations on the display.
Enclose value in double quotes (“)
when setting from the RS-232
interface.
CLOCK_ADJ
Sec./Day 0
-60–60
Time-of-day clock speed adjustment.
LANGUAGE_SELECT
— ENGL ENGL,
SECD,
EXTN
Selects the language to use for the
user interface. Enclose value in
double quotes (“) when setting from
the RS-232 interface.
MAINT_TIMEOUT
Hours
2
0.1–100
Time until automatically switching
out of software-controlled
maintenance mode.
MEASURE_PERIOD
Minutes
10
1–60
Length of time to measure each gas.
MEASURE_DELAY
Minutes
3
0.1–20
How long to defer sampling after
switching streams.
CONV_TIME
—
33 MS
33 MS,
66 MS,
133 MS,
266 MS,
533 MS,
1 SEC,
2 SEC
Conversion time for PMT and UV
detector channels. Enclose value in
double quotes (“) when setting from
the RS-232 interface.