Control inputs – Teledyne 6200E - Sulfides Analyzer User Manual

Page 140

background image

Operating Instructions

Model 6200E Instruction Manual

140

M6200E Rev: A1

Table 6-22: Status Output Pin Assignments

CONNECTOR

PIN

STATUS CONDITION

(ON=CONDUCTING)

1

System Ok

ON if no faults are present.

2

Conc Valid

ON if concentration measurement is valid, OFF when invalid.

3

High Range

ON if unit is in high range of any AUTO range mode.

4

Zero Cal

ON whenever the instrument is in ZERO calibration mode.

5

Span Cal

ON whenever the instrument is in SPAN calibration mode.

6

Diag Mode

ON whenever the instrument is in DIAGNOSTIC mode.

7

Low Range

ON if unit is in low range of any AUTO range mode.

8

Unused

D

Emitter Bus

The emitters of the transistors on pins 1-8 are bussed
together. For most applications, this pin should be connected
to the circuit ground of the receiving device.

+

Dc Power

+ 5 VDC source, 30 mA maximum (combined rating with
Control Inputs)

Digital Ground

The ground from the analyzer’s internal, 5/±15 VDC power
supply.

6.12.1.2. Control Inputs

Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes are provided
through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel. These are opto-

isolated, digital inputs that are activated when a 5 VDC signal from the “U” pin is connected to the
respective input pin.