Control inputs – Teledyne 6200E - Sulfides Analyzer User Manual
Page 140

Operating Instructions
Model 6200E Instruction Manual
140
M6200E Rev: A1
Table 6-22: Status Output Pin Assignments
CONNECTOR
PIN
STATUS CONDITION
(ON=CONDUCTING)
1
System Ok
ON if no faults are present.
2
Conc Valid
ON if concentration measurement is valid, OFF when invalid.
3
High Range
ON if unit is in high range of any AUTO range mode.
4
Zero Cal
ON whenever the instrument is in ZERO calibration mode.
5
Span Cal
ON whenever the instrument is in SPAN calibration mode.
6
Diag Mode
ON whenever the instrument is in DIAGNOSTIC mode.
7
Low Range
ON if unit is in low range of any AUTO range mode.
8
Unused
D
Emitter Bus
The emitters of the transistors on pins 1-8 are bussed
together. For most applications, this pin should be connected
to the circuit ground of the receiving device.
+
Dc Power
+ 5 VDC source, 30 mA maximum (combined rating with
Control Inputs)
Digital Ground
The ground from the analyzer’s internal, 5/±15 VDC power
supply.
6.12.1.2. Control Inputs
Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes are provided
through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel. These are opto-
isolated, digital inputs that are activated when a 5 VDC signal from the “U” pin is connected to the
respective input pin.