Operating Instructions

Model 6200E Instruction Manual

146

M6200E Rev: A1

To Initialize the modem press:


Select which

COM Port is

tested

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS DIAG ALRM

EXIT

SAMPLE

ENTER SETUP PASS : 818


8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU


CFG DAS

RNGE

PASS CLK

MORE

EXIT

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X


< TST TST > CAL

SETUP

SETUP X.X

COMMUNICATIONS MENU


ID

COM1

COM2

EXIT

EXIT returns

to the

previous

menu

SETUP X.X

COM1 MODEM INIT:AT Y

∅ &D∅ &H


SET>

EDIT

EXIT

SETUP X.X

COM1 INITIALIZE MODEM


SET> INIT

EXIT

SETUP X.X

COM1 MODE:0


SET>

EDIT

EXIT

SETUP X.X COM1 BAUD RATE:19200


SET>

EDIT

EXIT

EXIT returns to the

Communications Menu.

SETUP X.X

INITIALIZING MODEM


SET> INIT

EXIT

6.12.2.7. COM Port Password Security

In order to provide security for remote access of the M6200E, a LOGON feature can be enabled to
require a password before the instrument will accept commands. This is done by turning on the
SECURITY MODE (Section 6.10.8). Once the SECURITY MODE is enabled, the following items
apply.

• A password is required before the port will respond or pass on commands.

• If the port is inactive for one hour, it will automatically logoff, which can also be achieved

with the LOGOFF command.

• Three unsuccessful attempts to log on with an incorrect password will cause subsequent

logins to be disabled for 1 hour, even if the correct password is used.

• If not logged on, the only active command is the '?' request for the help screen.

• The following messages will be returned at logon:

o

LOGON SUCCESSFUL - Correct password given

o

LOGON FAILED - Password not given or incorrect