Model 6200E Instruction Manual

APPENDIX A-2: Setup Variables For Serial I/O, Revision A.1

M6200E Rev A.1

285

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT

VALUE

VALUE RANGE

DESCRIPTION

H2S_SPAN1

Conc 400

0.1–50000

Target

H

2

S concentration during span

calibration of range 1.

H2S_SPAN2

Conc 400

0.1–50000

Target

H

2

S concentration during span

calibration of range 2.

H2S_SLOPE1

PPB/mV 1

0.25–4

H

2

S slope for range 1.

H2S_SLOPE2

PPB/mV 1

0.25–4

H

2

S slope for range 2.

H2S_OFFSET1

mV 0

-1500–1500

H

2

S offset for range 1.

H2S_OFFSET2

mV 0

-1500–1500

H

2

S offset for range 2.

RANGE_MODE

— SNGL SNGL,

IND,
AUTO

Range control mode. Enclose value in

double quotes (“) when setting from
the RS-232 interface.

PHYS_RANGE1

PPM 2

0.1–2500

Low

pre-amp

range.

PHYS_RANGE2

PPM 20

0.1–2500

High

pre-amp

range.

CONC_RANGE1

Conc

500

0.1–50000

D/A concentration range 1.

CONC_RANGE2

Conc

500

0.1–50000

D/A concentration range 2.

SAMP_FLOW_SET

cc/m 700

Warnings
:
350–
1200

0–1200 Sample

flow

set point for flow

calculation and warning limits.

SAMP_FLOW_SLOPE

— 1

0.5–1.5

Sample

flow

slope correction factor

(adjusted flow = measured flow x

slope).

SAMP_PRESS_SET

"Hg 29.92

Warnings
:
15–35

0–100

Sample pressure set point for
pressure compensation and warning

limits.

SAMP_PRESS_SLOPE

— 1

0.5–1.5

Sample

pressure slope correction

factor (adjusted pressure =
measured pressure x slope).

CONV_TYPE

2

MOLY

NONE, MOLY

Converter type.

CONV_SET

2

ºC 315

Warnings
:
310–320

0–350

Converter temperature set point and
warning limits.

BOX_SET

ºC 30

Warnings
:
8–50

5–60

Box temperature warning limits. Set
point is not used.

PMT_SET

ºC 7

Warnings

:
2–12

0–40

PMT temperature set point and

warning limits.