Model 6200E Instruction Manual
APPENDIX A-2: Setup Variables For Serial I/O, Revision A.1
M6200E Rev A.1
285
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
H2S_SPAN1
Conc 400
0.1–50000
Target
H
2
S concentration during span
calibration of range 1.
H2S_SPAN2
Conc 400
0.1–50000
Target
H
2
S concentration during span
calibration of range 2.
H2S_SLOPE1
PPB/mV 1
0.25–4
H
2
S slope for range 1.
H2S_SLOPE2
PPB/mV 1
0.25–4
H
2
S slope for range 2.
H2S_OFFSET1
mV 0
-1500–1500
H
2
S offset for range 1.
H2S_OFFSET2
mV 0
-1500–1500
H
2
S offset for range 2.
RANGE_MODE
— SNGL SNGL,
IND,
AUTO
Range control mode. Enclose value in
double quotes (“) when setting from
the RS-232 interface.
PHYS_RANGE1
PPM 2
0.1–2500
Low
pre-amp
range.
PHYS_RANGE2
PPM 20
0.1–2500
High
pre-amp
range.
CONC_RANGE1
Conc
500
0.1–50000
D/A concentration range 1.
CONC_RANGE2
Conc
500
0.1–50000
D/A concentration range 2.
SAMP_FLOW_SET
cc/m 700
Warnings
:
350–
1200
0–1200 Sample
flow
set point for flow
calculation and warning limits.
SAMP_FLOW_SLOPE
— 1
0.5–1.5
Sample
flow
slope correction factor
(adjusted flow = measured flow x
slope).
SAMP_PRESS_SET
"Hg 29.92
Warnings
:
15–35
0–100
Sample pressure set point for
pressure compensation and warning
limits.
SAMP_PRESS_SLOPE
— 1
0.5–1.5
Sample
pressure slope correction
factor (adjusted pressure =
measured pressure x slope).
CONV_TYPE
2
—
MOLY
NONE, MOLY
Converter type.
CONV_SET
2
ºC 315
Warnings
:
310–320
0–350
Converter temperature set point and
warning limits.
BOX_SET
ºC 30
Warnings
:
8–50
5–60
Box temperature warning limits. Set
point is not used.
PMT_SET
ºC 7
Warnings
:
2–12
0–40
PMT temperature set point and
warning limits.