beautypg.com

Analog output voltages, Status outputs – Teledyne 6200E - Sulfides Analyzer User Manual

Page 256

background image

TROUBLESHOOTING & REPAIR

Model 6200E Instruction Manual

256

M6200E Rev: A1

11.5.8.2. Analog Output Voltages

To verify that the analog outputs are working properly, connect a voltmeter to the output in
question and perform an analog output step test as described in Section 6.9.3.

For each of the steps, taking into account any offset that may have been programmed into the
channel (Section 6.9.4.4), the output should be within 1% of the nominal value listed in the table

below except for the 0% step, which should be within 2-3 mV. If one or more of the steps is
outside of this range, a failure of one or both D/A converters and their associated circuitry on the

motherboard is likely.

Table 11-7: Analog Output Test Function - Nominal Values

FULL SCALE OUTPUT VOLTAGE

100MV

1V

5V

10V

STEP

%

NOMINAL OUTPUT VOLTAGE

1 0 0

mV 0 0 0

2 20

20

mV

0.2

1 2

3 40

40

mV

0.4

2 4

4 60

60

mV

0.6

3 6

5 80

80

mV

0.8

4 8

6 100

100

mV

1.0 5 10

11.5.8.3. Status Outputs

The procedure below can be used to test the Status outputs.

1. Connect a cable jumper between the “-“ pin and the “V” pin on the status output connector.

2. Connect a 1000 Ω resistor between the +5 V and the pin for the status output that is being

tested.

Table 11-8: Status Outputs Check Pin Out

PIN

(left to right)

STATUS

1 System

Ok

2 Conc

Valid

3 High

Range

4 Zero

Cal

5 Span

Cal

6 Diag

Mode

7 Spare
8 Spare

3. Connect a voltmeter between the “-“ pin and the pin of the output being tested (Table 11-8).

4. Under the

DIAG / SIGNAL I/O menu (Section 6.9.2), scroll through the inputs and outputs

until you get to the output in question. Alternately turn on and off the output noting the
voltage on the voltmeter, it should vary between 0 volts for ON and 5 volts for OFF.