beautypg.com

Teledyne 6200E - Sulfides Analyzer User Manual

Page 114

background image

Operating Instructions

Model 6200E Instruction Manual

114

M6200E Rev: A1


Each analyzer in the multidrop chain must have:

• One Teledyne Analytical Instruments option 62 installed.

• One 6’ straight-through, DB9 male Æ DB9 Female cable (Teledyne Analytical Instruments

P/N WR0000101) is required for each analyzer.

To set up the network, for each analyzer:

1. Turn the analyzer on and change its ID code (see Section 6.10.1) to a unique 4-digit number.

2. Remove the top cover (see Section 3.1) of the analyzer and locate JP2 on the multidrop PCA

(see Figure 6-12)

3. Make sure that the jumpers are in place connection pins 9

10 and 11

12.

4. If the analyzer is to be the last instrument on the chain, make sure a jumper is in place

connecting pins 21

22.

5. If you are adding an analyzer to the end of an already existing chain, don’t forget to remove

JP2, pins 21

22 on the multidrop PCA on the analyzer that was previous the last instrument

in the chain.

6. Close the instrument.

7. Using straight-through, DB9 male Æ DB9 Female cables, interconnect the host and the

analyzers as shown in Figure 6-13.

NOTE:

Teledyne Analytical Instruments recommends setting up the first link, between the Host

and the first analyzer and testing it before setting up the rest of the chain.