Range units, Mols per mol, ppb ), parts per million (10, Ug ) or milligrams per cubic meter (mg/m – Teledyne 6200E - Sulfides Analyzer User Manual

Page 76: And µg/m, Ppb x 2.86 = µg/m, Ppm x 2.86 = mg/m, S: ppb x 1.52 = µg/m, Ppm x 1.52 = mg/m

background image

Operating Instructions

Model 6200E Instruction Manual

76

M6200E Rev: A1

NOTE

On analyzers with the multigas option activated (see section 5.5 and Section 6.8.1) the

concentration value will switch back and forth between from “H2S=XXX.X” to

“SO2=XXX.X” depending on which gas is currently being measured.

ALSO

The analyzer will switch between the HIGH and LOW analog reporting ranges whenever

the concentration level of the gas being currently measured fulfills the trigger criteria

listed at the beginning of this section.

6.7.7. Range Units

The M6200E can display concentrations in parts per billion (10

9

mols per mol, PPB), parts per

million (10

6

mols per mol, PPM), micrograms per cubic meter (µg/m

3

, UG) or milligrams per cubic

meter (mg/m

3

, MG). Changing units affects all of the display, analog outputs, COM port and iDAS

values for all reporting ranges regardless of the analyzer’s range mode.

To change the concentration units:

SETUP X.X

CONC UNITS: PPM

PPM PPB

UGM

MGM %

ENTER EXIT

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK MORE

EXIT

SETUP X.X

RANGE CONTROL MENU

MODE

SET

UNIT

EXIT

SETUP X.X

CONC UNITS: PPB

PPM

PPB

UGM

MGM

ENTER

EXIT

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

Select the preferred

concentration unit.

EXIT returns

to the main menu.

ENTR accepts

the new unit,
EXIT returns

to the SETUP

menu.

NOTE

Concentrations displayed in mg/m

3

and µg/m

3

use standard temperature and pressure

(STP). The conversion factors from volumetric to mass units used in the M6200E are:

SO

2

: ppb x 2.86 = µg/m

3

; ppm x 2.86 = mg/m

3

H

2

S: ppb x 1.52 = µg/m

3

; ppm x 1.52 = mg/m

3