Range units, Mols per mol, ppb ), parts per million (10, Ug ) or milligrams per cubic meter (mg/m – Teledyne 6200E - Sulfides Analyzer User Manual
Page 76: And µg/m, Ppb x 2.86 = µg/m, Ppm x 2.86 = mg/m, S: ppb x 1.52 = µg/m, Ppm x 1.52 = mg/m
Operating Instructions
Model 6200E Instruction Manual
76
M6200E Rev: A1
NOTE
On analyzers with the multigas option activated (see section 5.5 and Section 6.8.1) the
concentration value will switch back and forth between from “H2S=XXX.X” to
“SO2=XXX.X” depending on which gas is currently being measured.
ALSO
The analyzer will switch between the HIGH and LOW analog reporting ranges whenever
the concentration level of the gas being currently measured fulfills the trigger criteria
listed at the beginning of this section.
6.7.7. Range Units
The M6200E can display concentrations in parts per billion (10
9
mols per mol, PPB), parts per
million (10
6
mols per mol, PPM), micrograms per cubic meter (µg/m
3
, UG) or milligrams per cubic
meter (mg/m
3
, MG). Changing units affects all of the display, analog outputs, COM port and iDAS
values for all reporting ranges regardless of the analyzer’s range mode.
To change the concentration units:
SETUP X.X
CONC UNITS: PPM
PPM PPB
UGM
MGM %
ENTER EXIT
SAMPLE RANGE = 500.0 PPB
H2S
=XXX.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
CONC UNITS: PPB
PPM
PPB
UGM
MGM
ENTER
EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
Select the preferred
concentration unit.
EXIT returns
to the main menu.
ENTR accepts
the new unit,
EXIT returns
to the SETUP
menu.
NOTE
Concentrations displayed in mg/m
3
and µg/m
3
use standard temperature and pressure
(STP). The conversion factors from volumetric to mass units used in the M6200E are:
SO
2
: ppb x 2.86 = µg/m
3
; ppm x 2.86 = mg/m
3
H
2
S: ppb x 1.52 = µg/m
3
; ppm x 1.52 = mg/m
3