Teledyne 6200E - Sulfides Analyzer User Manual
Page 111
Model 6200E Instruction Manual
Operating Instructions
M6200E Rev: A1
111
STEP 2: Configure the INSTRUMENT IP, GATEWAY IP and SUBNET MASK addresses by
pressing:
Internet Configuration Keypad Functions
KEY FUNCTION
[0]
Press this key to cycle through the range of
numerals and available characters (“0 – 9” & “ . ”)
Moves the cursor one character left or right.
DEL
Deletes a character at the cursor location.
ENTR
Accepts the new setting and returns to the previous
menu.
EXIT
Ignores the new setting and returns to the previous
menu.
Some keys only appear as needed.
SETUP X.X
DHCP: OFF
SET>
EDIT
EXIT
SETUP X.X
INST IP: 000.000.000.000
EXIT
SETUP X.X
GATEWAY IP: 000.000.000.000
EXIT
SETUP X.X
INST IP: [0] 00.000.000
ENTR EXIT
SETUP X.X
SUBNET MASK:255.255.255.0
EXIT
SETUP X.X
SUBNET MASK:[2]55.255.255.0
ENTR EXIT
SETUP X.X
TCP PORT 3000
EDIT
EXIT
The PORT number needs to remain at 3000.
Do not change this setting unless instructed to by
Teledyne Instruments Customer Service personnel.
From Step 1 above)
SETUP X.X
GATEWAY IP: [0] 00.000.000
ENTR EXIT
Cursor
location is
indicated by
brackets
SETUP X.X
INITIALIZING INET 0%
…
INITIALIZING INET 100%
SETUP X.X
INITIALIZATI0N SUCCEEDED
SETUP X.X
INITIALIZATION FAILED
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
EXIT
Pressing EXIT from
any of the above
display menus
causes the Ethernet
option to reinitialize
its internal interface
firmware
Contact your IT
Network Administrator