6.9.10. Test Channel Output

When activated, output channel A4 can be used to report one of the test functions viewable from

the SAMPLE mode display. To activate the A4 channel and select a test function, follow this key
sequence:

SAMPLE RANGE = 500.0 PPB

H2S

=XXX.X

< TST TST > CAL

SETUP

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

DIAG

SIGNAL I / O

NEXT

ENTR

EXIT

DIAG

ANALOG OUTPUT

PREV

NEXT

ENTR

EXIT

DIAG TCHN

TEST CHANNEL: NONE

NEXT

ENTR

EXIT

DIAG

TEST CHAN OUTPUT

PREV

NEXT

ENTR

EXIT

EXIT returns

to the main

SAMPLE

display

Continue to press NEXT until …

DIAG TCHN

TEST CHANNEL: PMT READING

PREV

NEXT

ENTR

EXIT

Press PREV or NEXT

to move through the

list of available

parameters

(Table 6-13)

Press EXIT to

return to the
DIAG menu

Press ENTR to select

the displayed

parameter activating

the test channel.

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

Table 6-14: Test Parameters Available for Analog Output A4

TEST CHANNEL

TEST PARAMETER RANGE

1

NONE

Test channel is turned off

PMT READING

0-5000 mV

UV READING

0-5000 mV

SAMPLE PRESSURE

0-40 in-Hg-A

SAMPLE FLOW

0-1000 cm³/min

RCELL TEMP

0-70° C

CHASSIS TEMP

0-70° C

IZS TEMP

0-70° C

PMT TEMP

0-50° C

CHASSIS TEMP

0-70° C

HVPS VOLTAGE

0-5000 V

1

This refers to the voltage range of the parameter and

not the output signal of the test channel.