Model 6200E Instruction Manual

Calibration Procedures

M6200E Rev: A1

171

7.7. Manual Calibration in INDEPENDENT or AUTO

Reporting Range Modes

If the analyzer is being operated in IND or AUTO Range mode, then the High and Low ranges
must be independently checked.

When the analyzer is in either Independent or Auto Range modes the user must run a separate
calibration procedure for each range. After pressing the CAL, CALZ or CALS keys the user is

prompted for the range that is to be calibrated as seen in the CALZ example below:

Continue Calibration as per

Standard Procedure

SAMPLE H2S STB = XXX.X PPB

H2S

=XXX.X

< TST TST > CAL

CALZ

CALS

SETUP

ZERO CAL M

H2S STB XXX.XX PPB

H2S

=XXX.X

< TST TST >

ZERO

CONC

EXIT

Analyzer enters

ZERO CAL Mode

See Table 5-1 for Z/S

Valve States during

this operating mode

SAMPLE

RANGE TO CAL: LOW

LOW

HIGH

ENTR SETUP

SAMPLE

RANGE TO CAL: HIGH

LOW HIGH

ENTR

SETUP

Wait until H2S

STB falls below

0.5 ppb. This may

take several

minutes.

Once this selection is made, the calibration procedure continues as previously described in Sections

7.2 through 7.6. The other range may be calibrated by starting over from the main SAMPLE
display.

7.7.1. Calibration With Remote Contact Closures

Contact closures for controlling calibration and calibration checks are located on the rear panel
CONTROL IN connector. Instructions for setup and use of these contacts can be found in Section
6.12.2.

When the appropriate contacts are closed for at least 5 seconds, the instrument switches into zero,
low span or high span mode and the internal zero/span valves will be automatically switched to the

appropriate configuration. The remote calibration contact closures may be activated in any order. It
is recommended that contact closures remain closed for at least 10 minutes to establish a reliable
reading; the instrument will stay in the selected mode for as long as the contacts remain closed.

If contact closures are used in conjunction with the analyzer’s AutoCal (Section 7.9) feature and
the AutoCal attribute CALIBRATE is enabled, the M6200E will not re-calibrate the analyzer until