beautypg.com

Teledyne 6200E - Sulfides Analyzer User Manual

Page 29

background image

Model 6200E Instruction Manual

Getting Started

M6200E Rev: A1

29

NOTE

Maximum pressure of any gas at the sample inlet should not exceed 1.5 in-Hg above

ambient pressure and ideally should equal ambient atmospheric pressure.

In applications where the sample gas is received from a pressurized manifold, a vent

must be provided to equalize the sample gas with ambient atmospheric pressure before

it enters the analyzer. The vented gas needs to be routed outside the immediate area or

shelter surrounding the instrument.

3. Attach zero air and span gas supply lines as appropriate (see Figures 3-6 & 3.7). For this type

of analyzer, zero air and span gas are defined as follows:

Zero air and span gas inlets should supply their respective gases in excess of the 700 cc

3

/min

demand of the analyzer. Supply and vent lines should be of sufficient length and diameter to

prevent back diffusion and pressure effects.

SPAN GAS

• A gas specifically mixed to match the chemical composition of the type of gas being

measured at near full scale of the desired measurement range. In the case of H

2

S,

measurements made with the Teledyne Analytical Instruments Model 6200E UV

Fluorescence H

2

S Analyzer it is recommended that you use a span gas with a H

2

S

concentration equal to 90% of the measurement range for your application.

EXAMPLE: If the application is to measure between 0 ppb and 500 ppb, an appropriate
span gas concentration would be 450 ppb H

2

S in N

2

.

Cylinders of calibrated H

2

S gas traceable to NIST-Standard Reference Material specifications

(also referred to as SRM’s or EPA protocol calibration gases) are commercially available. Table

3-5 lists specific NIST-SRM reference numbers for various concentrations of H

2

S.

Some applications, such as EPA monitoring, require a multipoint calibration procedure

where span gases of different concentrations are needed. We recommend using a bottle of

calibrated H

2

S gas of higher concentration in conjunction with a gas dilution calibrator such

as a Teledyne Analytical Instruments Model 700. This type of calibrator precisely mixes a

high concentration gas from zero air (both supplied externally) to accurately produce span

gas of the correct concentration. Linearity profiles can be automated with this model and

run unattended over night.

Table 3-5: NIST-SRM's Available for Traceability of H

2

S & SO

2

Calibration Gases

NIST-SRM

4

TYPE

NOMINAL

CONCENTRATION

2730
2731

Hydrogen sulfide in N

2

Hydrogen sulfide in N

2

5000 ppb

20 ppm

1693a

1694a
1661a

Sulfur dioxide in N

2

Sulfur dioxide in N

2

Sulfur dioxide in N

2

50 ppm

100 ppm
500 ppm